Study design and methods

Primary relevance to vitamin D Vitamin D is a stated subject of this paper.

Vitamin D is a stated subject of this paper; the supporting source evidence is shown below.

Why this paper is included

Supporting source evidence:

Effect of extracting solvent′s volume on the obtained responses from the analysis of vitamin D 3 .

Source: Results and Discussion. View original paper

Relevance classification does not establish study quality, causation or editorial approval.

Study design Study design awaiting verification.

Study design awaiting verification

The available source does not establish a sufficiently clear study design.

Meta-analysis unclear · Blinding unclear / not reported · Research setting awaiting verification

How this paper was classified

Automatic source-based classification; awaiting editorial verification. This is not a quality score.

View supporting evidence in the original paper

What the study types mean
Systematic review
Uses explicit methods to identify, assess and summarise research on a defined question.
Scoping review
Maps the extent and characteristics of research on a topic.
Narrative review
Discusses existing research; systematic-review methods have not been established by this classification.
Review / evidence synthesis
Reviews existing research; the exact review methods need further assessment.
Randomised controlled trial
Assigns participants or groups to comparison interventions by chance.
Non-randomised intervention study
Examines an intervention without random allocation.
Cohort study
Examines exposures and outcomes over time within a defined group.
Case-control study
Compares people with an outcome against people without it, examining exposures.
Cross-sectional study
Examines a population at a particular point in time.
Observational study
Observes exposures and outcomes without assigning the exposure; the subtype is not established here.
Case report
Describes an individual case.
Case series
Describes a group of individual cases.
Secondary analysis
Analyses existing study data; this paper is not automatically another independent trial.
Laboratory study
Investigates biological processes in a laboratory setting.
Qualitative study
Examines experiences or meanings using qualitative methods.
Genetic instrumental-variable study
Uses genetic variation as an instrument to investigate relationships; this is not a randomised treatment trial.
Study protocol
Describes planned methods and is not evidence that the planned study has reported results.
Multiple reported designs
More than one design is explicitly reported; examine the supporting methods.
Study design awaiting verification
The available source does not establish a sufficiently clear study design.

Meta-analysis combines numerical results from studies. Blinding describes who was unaware of treatment allocation. Neither is a separate quality grade. Study design, vitamin D relevance and document format are independent filters.

Full studies

A Dispersive Liquid-Liquid Micro-Extraction Technique for the Pre-concentration and Quantification of Vitamin D3 in Milk and Yogurt Samples Using a Non-Aqueous HPLC Method

Study design awaiting verification · Iranian Journal of Pharmaceutical Research : IJPR · 2019

Original publication title

A Dispersive Liquid: Liquid Micro: Extraction Technique for the Pre: concentration and Quantification of Vitamin D3 in Milk and Yogurt Samples Using a Non-Aqueous HPLC Method

Maryam Ghalebi; Elnaz Tamizi; Shirin Ahmadi; Ahad Sheikhloo; Mahboob Nemati

Original publication · Publisher record · CC BY 3.0

Source wording retained. Formatting, section navigation and display-title punctuation adapted for this research record.

Authors’ conclusion

A DLLME-HPLC-UV method has been developed and validated for the reliable extraction and determination of Vit D3 in dairy products, particularly milk and yogurt. The inherent advantages of DLLME procedure including simplicity and time and cost effectiveness, low amounts of required organic solvents and smaller sample volume along with lower LOQ value and shorter retention time (5.0 ± 0.8 min) are the main superiorities of the developed method to the previously reported ones. Nevertheless, it is worth saying that application of more sensitive detectors such as mass spectroscopy would result in the lower LOD and LOQ values as it can be seen in Table 1.

Finally, derived from the all above mentioned discussions, it can be concluded that the developed method could be considered as a promising technique in the monitoring of Vit D3 in dairy products and evaluation of the label claims of the fortified samples in food quality control laboratories.

An extract from the source publication. Read it in context.

At a glance

Study background

What background does the paper provide?

Cholecalciferol, vitamin D3 (Vit D3), is one of the isomers of vitamin D that is converted to its active form via hydroxylation in liver and kidney. It is of great importance in keeping normal calcium metabolism (1). In the other words, the main biological function of Vit D3 is to improve intestinal absorption of calcium and phosphate (2).

The required Vit D3 for the human body can be provided through either exposure of its precursor (7-dehydrocholesterol), existing in the epidermal layer of skin, to ultraviolet (UV) radiation or food intake (3). However, it should be mentioned that the amount of Vit D3 in food resources is little and only fishery products, particularly fatty fish such as salmon, mackerel, and also, herring and fish liver oils, are the good sources of Vit D3 (4).

In the case of dairy products, bovine milk is considered to be one of the Vit D3′s dietary origins, nevertheless the amount of Vit D3 in milk is too low and in the range of 4 to 40 IU/L (0.1 to 1.0 µg/L) (3). It has been demonstrated that Vit D3 deficiency is a common nutritional problem in all age groups in the world (5, 6). This deficiency could lead to the prevalence of rickets in children and osteomalacia in adults (7, 8).

Owing to the important role of vitamin D3 in human health, the development of analytical methods applicable to determine the accurate amounts of this micronutrient in food stuffs is necessary (9).

Since vitamins are present in low quantities and in combination with other compounds in foodstuffs, different techniques such as saponification, liquid: liquid extraction (LLE), and solid-phase extraction (SPE) have been utilized for the cleanup and extraction of Vit D3 from food samples (10, 11). However, some drawbacks such as use of large volumes of toxic solvents and being time-consuming make these techniques as costly procedures which are incompatible with the environment.

In recent years, significant efforts have been devoted to achieve simple alternative methods for sample preparation. Dispersive liquid: liquid microextraction (DLLME) is a relatively new technique introduced by Rezaee et al. in 2006 (12). It is an easy to operate, rapid and environmentally friendly technique (13). Low amounts of required organic solvents and fast sample preparation procedure are the main advantages of DLLME (14).

Literature review revealed that different analytical methods mainly based on high performance liquid chromatography (HPLC) have been reported for determination of Vit D3 in foodstuffs (3, 7, 10 and 15-20). Analytical characteristics such as shorter analysis time, higher specificity and selectivity, acceptable precision and reasonable accuracy make HPLC as a favorable method for the food analysis compared to the other methods (21). An overview of the reported methodologies for the determination of Vit D3 in food samples was shown in Table 1.

As it can be seen in this table, although DLLME technique was utilized for the extraction of Vit D3 from infant cereals, it has not been utilized for the extraction of Vit D3 in milk and dairy products so far (10). Therefore, since dairy products are accessible and affordable foods for different social classes that could be suitable carriers for Vit D3 fortification, the aim of the present study was optimization of a simple and sensitive DLLME technique coupled to HPLC with UV detection applicable to the quantification of vitamin D3 in some dairy products including milk and yogurt samples.

Table 1

The previously reported methods for the extraction and quantification of Vit D3 in foodstuffs

Sample Extraction procedure Analytical method Sample volume/weight Extraction
solvent's type and volume
LOQ
(ng/mL)
Retention time (min) Linear range (ng/mL) Recovery Reference
Fresh bovine milk
Commercial milk
SPELC: MS/MS10 mLHexane
50 mL
0.117.7-30 to 40%
Skimmed Milk
Whole milk
SaponificationHPLC15 mLDiethyl ether in
hexane
30 mL
2.616.020 to 100094 to 110%
Infant cerealsDLLMELC: DAD0.2-2 gCarbon tetrachloride
150 µL
1 to 10012.6-95 to 103%
Non-enriched milk
Enriched milk
LLEHPLC100 mLHexane
100 mL
-19.31 to 7.5-
Fortified infant formula
Milk
Milk powder
LLELC/MS4 mLIsooctane
10 mL
0.143.60-103.2%
Whole milkSPEHPLC100 gDiethyl ether: petroleum ether-20.7-95.4%
MilkSaponificationLC-MS/MS4 mLHexane: dichloromethane
12 mL
-17.7-98.9%
Fortified milkSaponification:
Extraction with hexane
UPLC: PDA20 mLHexane
8 mL
-11.92.5 to 6093%
Milk-based infant formulasNon-heating saponificationLC-MS0.5-2 g in 10 mL waterIsopropyl alcohol
10 mL
-10.71 to 10093 to 110%
Milk and yogurtDLLMEHPLC-UV1 mLCarbon tetrachloride
80 µL
252-6086.7-113.3%Present work
Source: Introduction Source: Introduction Source: Introduction Source: Introduction Source: Introduction Source: Introduction

Experimental

Apparatus

Analytical balance (A&D Weighing, San Jose, CA), vortex mixer (Scientific Industries Inc., Bohemia, NY), and high-speed centrifuge (Eppendorf, Hamburg, Germany) were used for sample preparation.

Chemicals and Reagents

Reagents

Chloroform, carbon tetrachloride, isopropyl alcohol, and ethanol were purchased from Merck (Darmstadt, Germany). Methanol and acetonitrile were obtained from Duksan Pure_Chemicals Co. (Gyeong gi-do, Korea). All solvents were of analytical or HPLC grade.

Standard Solutions

Cholecalciferol (Vit D3) was prepared from Sigma-Aldrich. Stock solution of Vit D3 with a concentration of 1000 μg/mL was freshly prepared via dissolving accurately weighed amount of cholecalciferol standard powder in chloroform. Working standard solutions of Vit D3, in a concentration range of 1 to 60 ng/mL, were prepared through diluting fresh stock solution using mobile phase (acetonitrile/methanol 90:10% v/v). All solutions were stored at 4 °C and covered with aluminum foil to be protected from light, prior to the analysis.

HPLC-UV Instrumentation

The chromatographic analyses were performed using a HPLC system (Knauer, Berlin, Germany). The detection was carried out using a UV detector (Knauer, Berlin, Germany) at 265 nm. The separation was achieved using a C8 column (250 mm × 4.6 mm, 5 μm) (Knauer, Berlin, Germany) at room temperature (25 °C). The mobile phase composed of a mixture of acetonitrile and methanol (90:10% v/v) used in isocratic elution mode at a flow rate of 1.2 mL/min.

Sampling

A total of 10 dairy products including 7 brands of milk and 3 brands of yogurt among available and well-known dairy brands in Tabriz, Iran were purchased from local supermarkets in a time period of May to July 2016. Among purchased products, four brands of milk samples (A, B, C, and E) and one brand of yogurt samples (D) were fortified with Vit D3. It should be mentioned that three samples of each brand with different batch numbers were prepared and analyzed in triplicate.

Sample Preparation/Clean up Procedure

One mL of milk samples and 1 mg of yogurt samples were shaken for 1min. Then, 2 mL of acetonitrile was added into the samples and vortex mixed for 2 min. After that, the samples were centrifuged for 10 min at 6000 rpm and finally the upper aqueous layer was transferred to another centrifuge tube for DLLME procedure.

DLLME Procedure

The applied DLLME technique was based on a previously reported work (10). However, some modifications were performed in reported technique to optimize the efficacy of the pre-concentration procedure for the intended purpose. In order to conduct DLLME, 80 μL of carbon tetrachloride as an extracting solvent was added to the obtained fraction from the previous step that was used as a dispersive solvent as well. After that the mixture was quickly injected into 5 mL of double distilled water using a syringe that resulted in the formation of a cloudy solution.

Then, the mixture was gently shaken for 1 min followed by centrifugation at 11000 rpm for 2 min that led to the sedimentation of the extracting solvent at the bottom of the conical tube. Finally, the sedimented phase was totally transferred to a microtube and evaporated to dryness at room temperature. The residue was reconstituted with 50 μL of acetonitrile and 20 μL of the obtained solution was injected into the HPLC.

Validation of Optimized DLLME-HPLC Method

The optimized DLLME-HPLC technique was validated for the quantitative purposes based on the ICH guidelines on validation of analytical procedures (22). In order to investigate linearity of the method and lower limit of quantitation (LLOQ) and upper limit of quantitation (ULOQ) amounts, the peak areas of calibration samples with concentrations in the range of 2 to 60 ng/mL calculated from the equation of ″the obtained peak area from calibration sample: peak area of blank sample″ were plotted against Vit D3 concentrations (n = 3 at each concentration).

For preparation of calibration samples, 100 µL of working standard solutions with concentrations in a range of 10 to 600 ng/mL were spiked into 1 mL of samples. To illustrate, 100 µL of a working standard solution with a concentration of 10 ng/mL was spiked into 1 mL of a milk sample to achieve a calibration sample with a concentration of 1 ng/mL.

LOD (limit of detection), of the method has been calculated using the equation of "LOD = 3.3 × σ/s", where σ is the standard deviation of peak areas from the analysis of six individual blank samples prepared based on the optimized conditions and s is the slope of the calibration equation.

The precision, including repeatability and intermediate precision, and accuracy of the method were investigated at 3 concentrations at the lower, the middle, and the upper levels of the calibration curve (at 2, 20 and 60 ng/mL Vit D3). Repeatability and intermediate precision were derived from the calculated amounts for each concentration using the calibration equation and peak areas obtained from the repeated analyses (n = 3) in one day and 3 days in a row, respectively. For accuracy determination, the samples with known concentrations of 2, 20, and 60 ng/mL Vit D3 were analyzed in triplicate, then the experimentally derived concentrations were calculated from the peak areas and calibration equation.

The accuracy of the method was presented as a relative error (RE%) calculated using the equation of ″((calculated concentration: nominal concentration)/nominal concentration) × 100″. The recovery of the method was evaluated at three concentration levels including 2, 10, and 40 ng/mL and reported as a percentage of the experimentally derived concentration to the nominal concentration.

Source: Experimental Source: Experimental Source: Experimental Source: Experimental Source: Experimental Source: Experimental Source: Experimental Source: Experimental Source: Experimental Source: Experimental Source: Experimental Source: Experimental Source: Experimental Source: Experimental Source: Experimental Source: Experimental Source: Experimental Source: Experimental Source: Experimental Source: Experimental

Results and Discussion

Optimization of the HPLC Condition

The chromatographic separation was optimized for the intended purpose. In order to select the stationary phase, performance of C8 and C18 (4.6 mm × 250 mm, 5 μm) columns for the separation of Vit D3 were investigated. Due to the poor peak shapes obtained using C18 column, C8 column was selected as the suitable stationary phase. To select the optimized mobile phase, different mixtures of solvent consisting of 5% acetonitrile in methanol, 10% acetonitrile in methanol, and 10% acetonitrile in isopropyl alcohol, were evaluated.

The obtained results indicated that the best separation was achieved using the mixture of acetonitrile and methanol with a ratio of 90:10% v/v with flow rate of 1.2 mL/min and an isocratic elution mode at 25 ºC. The obtained chromatograms from the HPLC-UV analysis can be seen in Figure 1.

Optimization of the DLLME Procedure

Selection of Extracting Solvent′s Type and Volume

Selection of a suitable extracting solvent is a critical factor that affects the efficiency of the DLLME procedure. Low solubility in water, lower density than water, good chromatographic behavior and high affinity to the analyte of interest are important criteria to choose a suitable extracting solvent (23). In the present study, tetrachloroethane, carbon tetrachloride, and chloroform were utilized as extracting solvents. The extractions were carried out with injecting 80 μL of each extracting solvent in combination with 2 mL of acetonitrile extract (as a disperser solvent) into the 5 mL water.

Considering the volume of sedimented phase and the required time for the evaporation of the applied solvent at the end of DLLME procedure, carbon tetrachloride was selected as the extracting solvent for the further analyses. In addition to the extracting solvent′s type, the impact of its volume on the efficacy of DLLME procedure was evaluated, where different volumes of carbon tetrachloride (80, 100, 150 and 200 μL) were utilized to accomplish the best extraction output.

As shown in Figure 2, 80 μL carbon tetrachloride was selected as an optimal extracting solvent, since in low volumes of extracting solvent (<80 μL) the two phase system was not formed.

Table 1

The previously reported methods for the extraction and quantification of Vit D3 in foodstuffs

Sample Extraction procedure Analytical method Sample volume/weight Extraction
solvent's type and volume
LOQ
(ng/mL)
Retention time (min) Linear range (ng/mL) Recovery Reference
Fresh bovine milk
Commercial milk
SPELC: MS/MS10 mLHexane
50 mL
0.117.7-30 to 40%
Skimmed Milk
Whole milk
SaponificationHPLC15 mLDiethyl ether in
hexane
30 mL
2.616.020 to 100094 to 110%
Infant cerealsDLLMELC: DAD0.2-2 gCarbon tetrachloride
150 µL
1 to 10012.6-95 to 103%
Non-enriched milk
Enriched milk
LLEHPLC100 mLHexane
100 mL
-19.31 to 7.5-
Fortified infant formula
Milk
Milk powder
LLELC/MS4 mLIsooctane
10 mL
0.143.60-103.2%
Whole milkSPEHPLC100 gDiethyl ether: petroleum ether-20.7-95.4%
MilkSaponificationLC-MS/MS4 mLHexane: dichloromethane
12 mL
-17.7-98.9%
Fortified milkSaponification:
Extraction with hexane
UPLC: PDA20 mLHexane
8 mL
-11.92.5 to 6093%
Milk-based infant formulasNon-heating saponificationLC-MS0.5-2 g in 10 mL waterIsopropyl alcohol
10 mL
-10.71 to 10093 to 110%
Milk and yogurtDLLMEHPLC-UV1 mLCarbon tetrachloride
80 µL
252-6086.7-113.3%Present work
Table 2

The obtained results from the accuracy and precision evaluations

Concentration (ng/mL) Accuracy (RE%) Repeatability (RSD%) Intermediate precision (RSD%)
2-1.511.47.2
20-2.17.94.4
60+0.63.91.2
Table 3

The obtained results from the quantification of Vit D 3 in studied real samples using the developed DLLME-HPLC-UV method

Real samples
(milk and yoghurt brands)
Label claim (IU * /mL) Vit D 3 (ng/mL) ± SD Vit D (IU * /mL) ± SD
3
A0.210.5 ± 1.50.42 ± 0.1
B_**12.4 ± 1.70.49 ± 0.1
C0.217.1 ± 1.20.68 ± 0.1
D0.514.3 ± 3.30.57 ± 0.0
E0.16.30 ± 0.90.25 ± 0.5
F_9.10 ± 2.70.36 ± 0.1
G_7.60 ± 1.30.30 ± 0.0
H_6.40 ± 1.80.25 ± 0.1
I_5.20 ± 0.80.21 ± 0.0
J_6.30 ± 0.20.25 ± 0.0
*

Each IU is equal to 25 ng Vit D 3 .

**

Not mentioned.

Figure 1

A dispersive liquid: liquid micro: extraction technique for the pre: concentration and quantification of vitamin D 3 in milk and yogurt samples using a non-aqueous HPLC method. The obtained chromatogram from the analysis of samples using a optimized HPLC method, separation conditions: C 8 column (250 mm × 4.6 mm, 5 μm), room temperature (25 °C), mobile phase of acetonitrile/methanol (90:10% v/v), flow rate of 1.2 mL/min, UV detection at 265 nm. (A) blank milk sample; (B) milk sample spiked with 50 ng/mL Vit D 3

Figure image at the original source: open the original paper. This file is not included in this text import.
Figure 2

A dispersive liquid: liquid micro: extraction technique for the pre: concentration and quantification of vitamin D 3 in milk and yogurt samples using a non-aqueous HPLC method. Effect of extracting solvent′s volume on the obtained responses from the analysis of vitamin D 3 . Experimental conditions: type and volume of disperser solvent: acetonitrile, 2 mL; Vit D 3 concentration: 20 µg/mL; HPLC conditions: C 8 (4.6 mm × 250 mm, 5 μm) column, acetonitrile/methanol (90:10% v/v) with flow rate of 1.2 mL/min as a mobile phase, isocratic elution mode at 25 ºC, UV detection at 265 nm.

Figure image at the original source: open the original paper. This file is not included in this text import.

Selection of Disperser Solvent′s Type and Volume

Miscibility of disperser solvent with both the organic phase (extracting solvent) and the sample solution is an essential factor in the selection of a suitable disperser solvent. It helps to disperse the extracting solvent in aqueous phase and as a result improves the extraction performance (23). In this study, acetonitrile, ethanol, and methanol were utilized as disperser solvents. According to the obtained results, application of acetonitrile as a disperser solvent resulted in the acceptable extraction efficiency. In addition to the disperser solvent′s type, its volume plays an important role in the performance of extraction procedure, because it could directly influence formation of cloudy solution between ternary component solvent systems.

Therefore, the impact of dispersive solvent’s volume on the extraction efficacy was evaluated, where distinctive volumes of acetonitrile (500, 1000, 2000 and 3000 μL) were utilized to conduct DLLME procedure. The observations revealed that at low volumes (>2000 μL), acetonitrile couldn′t appropriately scatter extraction solvent resulting in negligible formation of the cloudy solution. Additionally, it is worth saying that application of low volumes of acetonitrile could not effectively precipitate the proteins of the samples in the cleanup step. Higher volumes of acetonitrile (<2000 μL) resulted in increased solubility of analytes in the aqueous phase and consequently less extraction productivity.

Therefore, acetonitrile with a volume of 2000 μL was selected as the optimal dispersive solvent.

Selection of the Extraction Time

In DLLME, extraction time is characterized as a time period between the injection of the blend of a disperser solvent, extracting solvent and sample in the water and centrifugation (23).

In the present study, observations showed that alterations in the extraction time from 30 sec to 2 min did not affect the extraction efficiency, significantly suggesting the independency of the applied DLLME technique from extraction time. However, in order to keep the uniformity among the different analyses, 1 min was selected as an extraction time.

Method Validation Results

Method validation results indicated that the method was linear in the concentration range of 2 to 60 ng/mL with a calibration equation of y = 452385x: 526/73, where y was peak area and x was concentration of Vit D3 in µg/mL, and correlation coefficient of 0.9997. The LLOQ and ULOQ of the method was 2 and 60 ng/mL, respectively, owing to the possibility of the quantifications with acceptable precision (RSD% ≤ 20%) and accuracy (|RE%| ≤ 20%) at mentioned concentrations.

Since the obtained standard deviation among the responses of 6 blank samples was 136.0, the LOD of the method was calculated as 0.9 ng/mL.

The accuracy and precision results are reported in Table 2. As it can be seen in this table, the method was accurate (-2.1% ≤ RE% ≤ +0.6%) and precise (RSD% of less than 11.4%) enough for the quantification of Vit D3 in the linear range, and its recovery was in the acceptable range of 86.7 ± 9.5% to 113.3 ± 5.6%. It should also be noted that the recovery of vitamin D3 in yoghurt samples (D, E and G brands), by the proposed method was from 88.55%, 89.23, and 99.05%, respectively.

Real Sample Analysis

The samples purchased from local supermarkets as described in sampling section, were analyzed using the optimized DLLME-HPLC-UV method. As it can be followed in Table 3, the method successfully quantified the amount of Vit D3 in different samples and the findings indicated that in only one brand of fortified milk samples the obtained amount of Vit D3 was in accordance with the label claim; the finding highlights the need to consider stricter legislation for the monitoring of the claims on the labels of dairy products in Iran.

Source: Results and Discussion Source: Results and Discussion Source: Results and Discussion Source: Results and Discussion Source: Results and Discussion Source: Results and Discussion Source: Results and Discussion Source: Results and Discussion Source: Results and Discussion Source: Results and Discussion Source: Results and Discussion Source: Results and Discussion Source: Results and Discussion Source: Results and Discussion Source: Results and Discussion Source: Results and Discussion Source: Results and Discussion Source: Results and Discussion Source: Results and Discussion Source: Results and Discussion Source: Results and Discussion Source: Results and Discussion Source: Results and Discussion

Conclusion

A DLLME-HPLC-UV method has been developed and validated for the reliable extraction and determination of Vit D3 in dairy products, particularly milk and yogurt. The inherent advantages of DLLME procedure including simplicity and time and cost effectiveness, low amounts of required organic solvents and smaller sample volume along with lower LOQ value and shorter retention time (5.0 ± 0.8 min) are the main superiorities of the developed method to the previously reported ones. Nevertheless, it is worth saying that application of more sensitive detectors such as mass spectroscopy would result in the lower LOD and LOQ values as it can be seen in Table 1.

Finally, derived from the all above mentioned discussions, it can be concluded that the developed method could be considered as a promising technique in the monitoring of Vit D3 in dairy products and evaluation of the label claims of the fortified samples in food quality control laboratories.

Table 1

The previously reported methods for the extraction and quantification of Vit D3 in foodstuffs

Sample Extraction procedure Analytical method Sample volume/weight Extraction
solvent's type and volume
LOQ
(ng/mL)
Retention time (min) Linear range (ng/mL) Recovery Reference
Fresh bovine milk
Commercial milk
SPELC: MS/MS10 mLHexane
50 mL
0.117.7-30 to 40%
Skimmed Milk
Whole milk
SaponificationHPLC15 mLDiethyl ether in
hexane
30 mL
2.616.020 to 100094 to 110%
Infant cerealsDLLMELC: DAD0.2-2 gCarbon tetrachloride
150 µL
1 to 10012.6-95 to 103%
Non-enriched milk
Enriched milk
LLEHPLC100 mLHexane
100 mL
-19.31 to 7.5-
Fortified infant formula
Milk
Milk powder
LLELC/MS4 mLIsooctane
10 mL
0.143.60-103.2%
Whole milkSPEHPLC100 gDiethyl ether: petroleum ether-20.7-95.4%
MilkSaponificationLC-MS/MS4 mLHexane: dichloromethane
12 mL
-17.7-98.9%
Fortified milkSaponification:
Extraction with hexane
UPLC: PDA20 mLHexane
8 mL
-11.92.5 to 6093%
Milk-based infant formulasNon-heating saponificationLC-MS0.5-2 g in 10 mL waterIsopropyl alcohol
10 mL
-10.71 to 10093 to 110%
Milk and yogurtDLLMEHPLC-UV1 mLCarbon tetrachloride
80 µL
252-6086.7-113.3%Present work
Source: Conclusion Source: Conclusion
Authors’ abstract

In present study, a DLLME-HPLC-UV method was developed and validated for the extraction, pre: concentration, and subsequently quantification of vitamin D3 (Vit D3) in milk and yogurt samples. In order to be able to extract Vit D3 from studied samples efficiently, the DLLME procedure was optimized with respect to the parameters affecting the extraction efficacy, where acetonitrile (2 mL as disperser solvent) resulting from the protein precipitation procedure was mixed with 80 µL carbon tetrachloride (as an extraction solvent) respectively.

The extracted samples were quantitatively analyzed with a HPLC technique using a C8 column (250 mm × 4.6 mm, 5 μm) at room temperature (25 °C), mobile phase of acetonitrile/methanol (90:10% v/v) in isocratic elution mode at a flow rate of 1.2 mL/min and UV detection at 265 nm. The method validation results revealed that the method was linear in the concentration range of 2 to 60 ng/mL (r = 0.9997) with a LOD of 0.9 ng/mL and LLOQ of 2 ng/mL; the method was accurate (-2.1% ≤ RE% ≤ +0.6%) and precise (1.2% ≤RSD% ≤ 11.3%) and its recovery was in the range of 86.6 to 113.3%.

The obtained results indicated that the method could be utilized as an easy to use technique for the monitoring Vit D3 in dairy products, especially milk and yogurt samples.

Read the full paper

Introduction

Cholecalciferol, vitamin D3 (Vit D3), is one of the isomers of vitamin D that is converted to its active form via hydroxylation in liver and kidney. It is of great importance in keeping normal calcium metabolism (1). In the other words, the main biological function of Vit D3 is to improve intestinal absorption of calcium and phosphate (2).

The required Vit D3 for the human body can be provided through either exposure of its precursor (7-dehydrocholesterol), existing in the epidermal layer of skin, to ultraviolet (UV) radiation or food intake (3). However, it should be mentioned that the amount of Vit D3 in food resources is little and only fishery products, particularly fatty fish such as salmon, mackerel, and also, herring and fish liver oils, are the good sources of Vit D3 (4).

In the case of dairy products, bovine milk is considered to be one of the Vit D3′s dietary origins, nevertheless the amount of Vit D3 in milk is too low and in the range of 4 to 40 IU/L (0.1 to 1.0 µg/L) (3). It has been demonstrated that Vit D3 deficiency is a common nutritional problem in all age groups in the world (5, 6). This deficiency could lead to the prevalence of rickets in children and osteomalacia in adults (7, 8).

Owing to the important role of vitamin D3 in human health, the development of analytical methods applicable to determine the accurate amounts of this micronutrient in food stuffs is necessary (9).

Since vitamins are present in low quantities and in combination with other compounds in foodstuffs, different techniques such as saponification, liquid: liquid extraction (LLE), and solid-phase extraction (SPE) have been utilized for the cleanup and extraction of Vit D3 from food samples (10, 11). However, some drawbacks such as use of large volumes of toxic solvents and being time-consuming make these techniques as costly procedures which are incompatible with the environment.

In recent years, significant efforts have been devoted to achieve simple alternative methods for sample preparation. Dispersive liquid: liquid microextraction (DLLME) is a relatively new technique introduced by Rezaee et al. in 2006 (12). It is an easy to operate, rapid and environmentally friendly technique (13). Low amounts of required organic solvents and fast sample preparation procedure are the main advantages of DLLME (14).

Literature review revealed that different analytical methods mainly based on high performance liquid chromatography (HPLC) have been reported for determination of Vit D3 in foodstuffs (3, 7, 10 and 15-20). Analytical characteristics such as shorter analysis time, higher specificity and selectivity, acceptable precision and reasonable accuracy make HPLC as a favorable method for the food analysis compared to the other methods (21). An overview of the reported methodologies for the determination of Vit D3 in food samples was shown in Table 1.

As it can be seen in this table, although DLLME technique was utilized for the extraction of Vit D3 from infant cereals, it has not been utilized for the extraction of Vit D3 in milk and dairy products so far (10). Therefore, since dairy products are accessible and affordable foods for different social classes that could be suitable carriers for Vit D3 fortification, the aim of the present study was optimization of a simple and sensitive DLLME technique coupled to HPLC with UV detection applicable to the quantification of vitamin D3 in some dairy products including milk and yogurt samples.

Experimental

Apparatus

Analytical balance (A&D Weighing, San Jose, CA), vortex mixer (Scientific Industries Inc., Bohemia, NY), and high-speed centrifuge (Eppendorf, Hamburg, Germany) were used for sample preparation.

Chemicals and Reagents

Reagents

Chloroform, carbon tetrachloride, isopropyl alcohol, and ethanol were purchased from Merck (Darmstadt, Germany). Methanol and acetonitrile were obtained from Duksan Pure_Chemicals Co. (Gyeong gi-do, Korea). All solvents were of analytical or HPLC grade.

Standard Solutions

Cholecalciferol (Vit D3) was prepared from Sigma-Aldrich. Stock solution of Vit D3 with a concentration of 1000 μg/mL was freshly prepared via dissolving accurately weighed amount of cholecalciferol standard powder in chloroform. Working standard solutions of Vit D3, in a concentration range of 1 to 60 ng/mL, were prepared through diluting fresh stock solution using mobile phase (acetonitrile/methanol 90:10% v/v). All solutions were stored at 4 °C and covered with aluminum foil to be protected from light, prior to the analysis.

HPLC-UV Instrumentation

The chromatographic analyses were performed using a HPLC system (Knauer, Berlin, Germany). The detection was carried out using a UV detector (Knauer, Berlin, Germany) at 265 nm. The separation was achieved using a C8 column (250 mm × 4.6 mm, 5 μm) (Knauer, Berlin, Germany) at room temperature (25 °C). The mobile phase composed of a mixture of acetonitrile and methanol (90:10% v/v) used in isocratic elution mode at a flow rate of 1.2 mL/min.

Sampling

A total of 10 dairy products including 7 brands of milk and 3 brands of yogurt among available and well-known dairy brands in Tabriz, Iran were purchased from local supermarkets in a time period of May to July 2016. Among purchased products, four brands of milk samples (A, B, C, and E) and one brand of yogurt samples (D) were fortified with Vit D3. It should be mentioned that three samples of each brand with different batch numbers were prepared and analyzed in triplicate.

Sample Preparation/Clean up Procedure

One mL of milk samples and 1 mg of yogurt samples were shaken for 1min. Then, 2 mL of acetonitrile was added into the samples and vortex mixed for 2 min. After that, the samples were centrifuged for 10 min at 6000 rpm and finally the upper aqueous layer was transferred to another centrifuge tube for DLLME procedure.

DLLME Procedure

The applied DLLME technique was based on a previously reported work (10). However, some modifications were performed in reported technique to optimize the efficacy of the pre-concentration procedure for the intended purpose. In order to conduct DLLME, 80 μL of carbon tetrachloride as an extracting solvent was added to the obtained fraction from the previous step that was used as a dispersive solvent as well. After that the mixture was quickly injected into 5 mL of double distilled water using a syringe that resulted in the formation of a cloudy solution.

Then, the mixture was gently shaken for 1 min followed by centrifugation at 11000 rpm for 2 min that led to the sedimentation of the extracting solvent at the bottom of the conical tube. Finally, the sedimented phase was totally transferred to a microtube and evaporated to dryness at room temperature. The residue was reconstituted with 50 μL of acetonitrile and 20 μL of the obtained solution was injected into the HPLC.

Validation of Optimized DLLME-HPLC Method

The optimized DLLME-HPLC technique was validated for the quantitative purposes based on the ICH guidelines on validation of analytical procedures (22). In order to investigate linearity of the method and lower limit of quantitation (LLOQ) and upper limit of quantitation (ULOQ) amounts, the peak areas of calibration samples with concentrations in the range of 2 to 60 ng/mL calculated from the equation of ″the obtained peak area from calibration sample: peak area of blank sample″ were plotted against Vit D3 concentrations (n = 3 at each concentration).

For preparation of calibration samples, 100 µL of working standard solutions with concentrations in a range of 10 to 600 ng/mL were spiked into 1 mL of samples. To illustrate, 100 µL of a working standard solution with a concentration of 10 ng/mL was spiked into 1 mL of a milk sample to achieve a calibration sample with a concentration of 1 ng/mL.

LOD (limit of detection), of the method has been calculated using the equation of "LOD = 3.3 × σ/s", where σ is the standard deviation of peak areas from the analysis of six individual blank samples prepared based on the optimized conditions and s is the slope of the calibration equation.

The precision, including repeatability and intermediate precision, and accuracy of the method were investigated at 3 concentrations at the lower, the middle, and the upper levels of the calibration curve (at 2, 20 and 60 ng/mL Vit D3). Repeatability and intermediate precision were derived from the calculated amounts for each concentration using the calibration equation and peak areas obtained from the repeated analyses (n = 3) in one day and 3 days in a row, respectively. For accuracy determination, the samples with known concentrations of 2, 20, and 60 ng/mL Vit D3 were analyzed in triplicate, then the experimentally derived concentrations were calculated from the peak areas and calibration equation.

The accuracy of the method was presented as a relative error (RE%) calculated using the equation of ″((calculated concentration: nominal concentration)/nominal concentration) × 100″. The recovery of the method was evaluated at three concentration levels including 2, 10, and 40 ng/mL and reported as a percentage of the experimentally derived concentration to the nominal concentration.

Results and Discussion

Optimization of the HPLC Condition

The chromatographic separation was optimized for the intended purpose. In order to select the stationary phase, performance of C8 and C18 (4.6 mm × 250 mm, 5 μm) columns for the separation of Vit D3 were investigated. Due to the poor peak shapes obtained using C18 column, C8 column was selected as the suitable stationary phase. To select the optimized mobile phase, different mixtures of solvent consisting of 5% acetonitrile in methanol, 10% acetonitrile in methanol, and 10% acetonitrile in isopropyl alcohol, were evaluated.

The obtained results indicated that the best separation was achieved using the mixture of acetonitrile and methanol with a ratio of 90:10% v/v with flow rate of 1.2 mL/min and an isocratic elution mode at 25 ºC. The obtained chromatograms from the HPLC-UV analysis can be seen in Figure 1.

Optimization of the DLLME Procedure

Selection of Extracting Solvent′s Type and Volume

Selection of a suitable extracting solvent is a critical factor that affects the efficiency of the DLLME procedure. Low solubility in water, lower density than water, good chromatographic behavior and high affinity to the analyte of interest are important criteria to choose a suitable extracting solvent (23). In the present study, tetrachloroethane, carbon tetrachloride, and chloroform were utilized as extracting solvents. The extractions were carried out with injecting 80 μL of each extracting solvent in combination with 2 mL of acetonitrile extract (as a disperser solvent) into the 5 mL water.

Considering the volume of sedimented phase and the required time for the evaporation of the applied solvent at the end of DLLME procedure, carbon tetrachloride was selected as the extracting solvent for the further analyses. In addition to the extracting solvent′s type, the impact of its volume on the efficacy of DLLME procedure was evaluated, where different volumes of carbon tetrachloride (80, 100, 150 and 200 μL) were utilized to accomplish the best extraction output.

As shown in Figure 2, 80 μL carbon tetrachloride was selected as an optimal extracting solvent, since in low volumes of extracting solvent (<80 μL) the two phase system was not formed.

Table 1

The previously reported methods for the extraction and quantification of Vit D3 in foodstuffs

Sample Extraction procedure Analytical method Sample volume/weight Extraction
solvent's type and volume
LOQ
(ng/mL)
Retention time (min) Linear range (ng/mL) Recovery Reference
Fresh bovine milk
Commercial milk
SPELC: MS/MS10 mLHexane
50 mL
0.117.7-30 to 40%
Skimmed Milk
Whole milk
SaponificationHPLC15 mLDiethyl ether in
hexane
30 mL
2.616.020 to 100094 to 110%
Infant cerealsDLLMELC: DAD0.2-2 gCarbon tetrachloride
150 µL
1 to 10012.6-95 to 103%
Non-enriched milk
Enriched milk
LLEHPLC100 mLHexane
100 mL
-19.31 to 7.5-
Fortified infant formula
Milk
Milk powder
LLELC/MS4 mLIsooctane
10 mL
0.143.60-103.2%
Whole milkSPEHPLC100 gDiethyl ether: petroleum ether-20.7-95.4%
MilkSaponificationLC-MS/MS4 mLHexane: dichloromethane
12 mL
-17.7-98.9%
Fortified milkSaponification:
Extraction with hexane
UPLC: PDA20 mLHexane
8 mL
-11.92.5 to 6093%
Milk-based infant formulasNon-heating saponificationLC-MS0.5-2 g in 10 mL waterIsopropyl alcohol
10 mL
-10.71 to 10093 to 110%
Milk and yogurtDLLMEHPLC-UV1 mLCarbon tetrachloride
80 µL
252-6086.7-113.3%Present work
Table 2

The obtained results from the accuracy and precision evaluations

Concentration (ng/mL) Accuracy (RE%) Repeatability (RSD%) Intermediate precision (RSD%)
2-1.511.47.2
20-2.17.94.4
60+0.63.91.2
Table 3

The obtained results from the quantification of Vit D 3 in studied real samples using the developed DLLME-HPLC-UV method

Real samples
(milk and yoghurt brands)
Label claim (IU * /mL) Vit D 3 (ng/mL) ± SD Vit D (IU * /mL) ± SD
3
A0.210.5 ± 1.50.42 ± 0.1
B_**12.4 ± 1.70.49 ± 0.1
C0.217.1 ± 1.20.68 ± 0.1
D0.514.3 ± 3.30.57 ± 0.0
E0.16.30 ± 0.90.25 ± 0.5
F_9.10 ± 2.70.36 ± 0.1
G_7.60 ± 1.30.30 ± 0.0
H_6.40 ± 1.80.25 ± 0.1
I_5.20 ± 0.80.21 ± 0.0
J_6.30 ± 0.20.25 ± 0.0
*

Each IU is equal to 25 ng Vit D 3 .

**

Not mentioned.

Figure 1

A dispersive liquid: liquid micro: extraction technique for the pre: concentration and quantification of vitamin D 3 in milk and yogurt samples using a non-aqueous HPLC method. The obtained chromatogram from the analysis of samples using a optimized HPLC method, separation conditions: C 8 column (250 mm × 4.6 mm, 5 μm), room temperature (25 °C), mobile phase of acetonitrile/methanol (90:10% v/v), flow rate of 1.2 mL/min, UV detection at 265 nm. (A) blank milk sample; (B) milk sample spiked with 50 ng/mL Vit D 3

Figure image at the original source: open the original paper. This file is not included in this text import.
Figure 2

A dispersive liquid: liquid micro: extraction technique for the pre: concentration and quantification of vitamin D 3 in milk and yogurt samples using a non-aqueous HPLC method. Effect of extracting solvent′s volume on the obtained responses from the analysis of vitamin D 3 . Experimental conditions: type and volume of disperser solvent: acetonitrile, 2 mL; Vit D 3 concentration: 20 µg/mL; HPLC conditions: C 8 (4.6 mm × 250 mm, 5 μm) column, acetonitrile/methanol (90:10% v/v) with flow rate of 1.2 mL/min as a mobile phase, isocratic elution mode at 25 ºC, UV detection at 265 nm.

Figure image at the original source: open the original paper. This file is not included in this text import.

Selection of Disperser Solvent′s Type and Volume

Miscibility of disperser solvent with both the organic phase (extracting solvent) and the sample solution is an essential factor in the selection of a suitable disperser solvent. It helps to disperse the extracting solvent in aqueous phase and as a result improves the extraction performance (23). In this study, acetonitrile, ethanol, and methanol were utilized as disperser solvents. According to the obtained results, application of acetonitrile as a disperser solvent resulted in the acceptable extraction efficiency. In addition to the disperser solvent′s type, its volume plays an important role in the performance of extraction procedure, because it could directly influence formation of cloudy solution between ternary component solvent systems.

Therefore, the impact of dispersive solvent’s volume on the extraction efficacy was evaluated, where distinctive volumes of acetonitrile (500, 1000, 2000 and 3000 μL) were utilized to conduct DLLME procedure. The observations revealed that at low volumes (>2000 μL), acetonitrile couldn′t appropriately scatter extraction solvent resulting in negligible formation of the cloudy solution. Additionally, it is worth saying that application of low volumes of acetonitrile could not effectively precipitate the proteins of the samples in the cleanup step. Higher volumes of acetonitrile (<2000 μL) resulted in increased solubility of analytes in the aqueous phase and consequently less extraction productivity.

Therefore, acetonitrile with a volume of 2000 μL was selected as the optimal dispersive solvent.

Selection of the Extraction Time

In DLLME, extraction time is characterized as a time period between the injection of the blend of a disperser solvent, extracting solvent and sample in the water and centrifugation (23).

In the present study, observations showed that alterations in the extraction time from 30 sec to 2 min did not affect the extraction efficiency, significantly suggesting the independency of the applied DLLME technique from extraction time. However, in order to keep the uniformity among the different analyses, 1 min was selected as an extraction time.

Method Validation Results

Method validation results indicated that the method was linear in the concentration range of 2 to 60 ng/mL with a calibration equation of y = 452385x: 526/73, where y was peak area and x was concentration of Vit D3 in µg/mL, and correlation coefficient of 0.9997. The LLOQ and ULOQ of the method was 2 and 60 ng/mL, respectively, owing to the possibility of the quantifications with acceptable precision (RSD% ≤ 20%) and accuracy (|RE%| ≤ 20%) at mentioned concentrations.

Since the obtained standard deviation among the responses of 6 blank samples was 136.0, the LOD of the method was calculated as 0.9 ng/mL.

The accuracy and precision results are reported in Table 2. As it can be seen in this table, the method was accurate (-2.1% ≤ RE% ≤ +0.6%) and precise (RSD% of less than 11.4%) enough for the quantification of Vit D3 in the linear range, and its recovery was in the acceptable range of 86.7 ± 9.5% to 113.3 ± 5.6%. It should also be noted that the recovery of vitamin D3 in yoghurt samples (D, E and G brands), by the proposed method was from 88.55%, 89.23, and 99.05%, respectively.

Real Sample Analysis

The samples purchased from local supermarkets as described in sampling section, were analyzed using the optimized DLLME-HPLC-UV method. As it can be followed in Table 3, the method successfully quantified the amount of Vit D3 in different samples and the findings indicated that in only one brand of fortified milk samples the obtained amount of Vit D3 was in accordance with the label claim; the finding highlights the need to consider stricter legislation for the monitoring of the claims on the labels of dairy products in Iran.

Conclusion

A DLLME-HPLC-UV method has been developed and validated for the reliable extraction and determination of Vit D3 in dairy products, particularly milk and yogurt. The inherent advantages of DLLME procedure including simplicity and time and cost effectiveness, low amounts of required organic solvents and smaller sample volume along with lower LOQ value and shorter retention time (5.0 ± 0.8 min) are the main superiorities of the developed method to the previously reported ones. Nevertheless, it is worth saying that application of more sensitive detectors such as mass spectroscopy would result in the lower LOD and LOQ values as it can be seen in Table 1.

Finally, derived from the all above mentioned discussions, it can be concluded that the developed method could be considered as a promising technique in the monitoring of Vit D3 in dairy products and evaluation of the label claims of the fortified samples in food quality control laboratories.

Acknowledgment

This work is a part of thesis by Maryam Ghalebi submitted for MSc degree (No. 4) and is supported by the Research Council, Tabriz University of Medical Sciences.

References

1 Kwak BM Jeong IS Lee MS Ahn JH Park JS Rapid determination of vitamin D3 in milk-based infant formulas by liquid chromatography-tandem mass spectrometry Food Chem 2014 165 569 74 25038713 10.1016/j.foodchem.2014.05.137
2 Bilodeau L Dufresne G Deeks J Clément G Bertrand J Turcotte S Robichaud A Beraldin F Fouquet A Determination of vitamin D3 and 25-hydroxyvitamin D3 in foodstuffs by HPLC UV-DAD and LC: MS/MS Food Comp 2011 24 441 8
3 Trenerry VC Plozza T Caridi D Murphy S The determination of vitamin D3 in bovine milk by liquid chromatography mass spectrometry Food Chem 2011 125 1314 9
4 Phillips K M Byrdwell WC Exler J Gebhardt SE Harnly JM Holden J M Horst RL Linda EL Patterson KY Wolf WR Development and validation of control materials for the measurement of vitamin D3 in selected US foods J. Food Comp. Anal 2011 24 299 306
5 Black LJ Seamans K M Cashman KD Kiely M An updated systematic review and meta-analysis of the efficacy of vitamin D food fortification J. Nutr 2012 142 1102 8 22513988 10.3945/jn.112.158014
6 Omidvar N Abtahi M Neyestani T Hajifaraji M Sensory evaluation and assessment of compliance with vitamin D and Calcium fortified-milk among school-age children in Tehran Iran. J. Nutr. Sci. Food Technol 2012 7 61 7
7 Chen Y Reddy MR Li W Yettlla RR Lopez S Development and validation of a high performance liquid chromatographic method for simultaneous determination of Vitamins A and D3 in fluid milk products J. AOAC Int 2015 98 390 6 25905745 10.5740/jaoacint.14-137
8 Kiely M Black LJ Dietary strategies to maintain adequacy of circulating 25-hydroxyvitamin D concentrations J. Scand. J. Clin. Lab. Inves 2012 72 14 23 10.3109/00365513.2012.681893 [link unavailable; original address: https://doi.org/10.3109/00365513.2012.681893] 22536758
9 Kasalova E Aufartova J Kujovska L mova K Solichova D Solich P Recent trends in the analysis of vitamin D and its metabolites in milk: A review Food Chem 2015 171 177 90 25308658 10.1016/j.foodchem.2014.08.102
10 Viñas P Bravo-Bravo M López-García I Hernández-Córdoba M Dispersive liquid: liquid microextraction for the determination of vitamins D and K in foods by liquid chromatography with diode-array and atmospheric pressure chemical ionization-mass spectrometry detection Talanta 2013 11 806 13 10.1016/j.talanta.2013.06.050 24054666
11 Rezaee M Yamini Y Faraji M Evolution of dispersive liquid: liquid microextraction method J. Chromatogr. A 2010 1217 2342 57 20005521 10.1016/j.chroma.2009.11.088
12 Rezaee M Assadi Y Milani MR Aghaee E Ahmadi F Berijani S Determination of organic compounds in water using dispersive liquid: liquid microextraction J. Chromatogr. A 2006 1116 1 9 16574135 10.1016/j.chroma.2006.03.007
13 Viñas P Bravo-Bravo M López-García I Pastor-Belda M Hernández-Córdoba M Pressurized liquid extraction and dispersive liquid: liquid microextraction for determination of tocopherols and tocotrienols in plant foods by liquid chromatography with fluorescence and atmospheric pressure chemical ionization-mass spectrometry detection Talanta 2014 119 98 104 24401390 10.1016/j.talanta.2013.10.053
14 Viñas P Bravo-Bravo M López-García I Hernández-Córdoba M Quantification of β-carotene, retinol, retinyl acetate and retinyl palmitate in enriched fruit juices using dispersive liquid: liquid microextraction coupled to liquid chromatography with fluorescence detection and atmospheric pressure chemical ionization-mass spectrometry J. Chromatogr. A 2013 1275 1 8 23290361 10.1016/j.chroma.2012.12.022
15 Barba FJ Esteve MJ Frígola A Determination of vitamins E (α-, γ-and δ-tocopherol) and D (cholecalciferol and ergocalciferol) by liquid chromatography in milk, fruit juice and vegetable beverage Eur. Food Res. Technol 2011 232 829 36
16 Abernethy GA A rapid analytical method for cholecalciferol (vitamin D3) in fortified infant formula, milk and milk powder using Diels: Alder derivatisation and liquid chromatography: tandem mass spectrometric detection Anal. Bioanal. Chem 2012 403 1433 40 22441200 10.1007/s00216-012-5939-1
17 Jakobsen J Saxholt E Vitamin D metabolites in bovine milk and butter J. Food Comp. Anal 2009 22 472 8
18 Gomes FP Shaw PN Whitfield K Hewavitharana AK Simultaneous quantitative analysis of eight vitamin D analogues in milk using liquid chromatography: tandem mass spectrometry Anal. Chim. Acta 2015 891 211 20 26388380 10.1016/j.aca.2015.08.017
19 Alvi SN El Tabache C Hammami MM Determination of vitamin-D level in fortified milk by ultra-performance liquid chromatography WorldJ. Pharm. Pharm. Sci 2015 4 170 81
20 Kwak BM Jeong IS Lee MS Ahn JH Park JS Rapid determination of vitamin D3 in milk-based infant formulas by liquid chromatography-tandem mass spectrometry Food Chem 2014 165 569 74 25038713 10.1016/j.foodchem.2014.05.137
21 Nollet LML Food Analysis by HPLC 2013 3nd edNew YorkTaylor & Francis330
22 ICH International conference on harmonisation of technical requirements for registration of pharmacuticals for human use, Q2 (r1): validation of analytical procedures: text and methodology 2005
23 Shammugasamy B Ramakrishnan Y Ghazali HM Muhammad K Combination of saponification and dispersive liquid: liquid microextraction for the determination of tocopherols and tocotrienols in cereals by reversed-phase high-performance liquid chromatography J. Chromatogr. A 2013 1300 31 7 23587317 10.1016/j.chroma.2013.03.036
Publication and source details

Original title: A Dispersive Liquid: Liquid Micro: Extraction Technique for the Pre: concentration and Quantification of Vitamin D3 in Milk and Yogurt Samples Using a Non-Aqueous HPLC Method

Maryam Ghalebi; Elnaz Tamizi; Shirin Ahmadi; Ahad Sheikhloo; Mahboob Nemati

Iranian Journal of Pharmaceutical Research : IJPR · 2019

* Corresponding author: E-mail: nematim@tbzmed.ac.irhttps://creativecommons.org/licenses/by/3.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution License, (http://creativecommons.org/licenses/by/3.0/) which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Repository JATS snapshot; version as supplied

Source retrieved: 2026-11-11T19:27:09.029271+00:00

DOI: 10.22037/ijpr.2019.1100634 · PMC6706729