Secondary relevance to vitamin D This paper mainly addresses another topic and also reports a vitamin D analysis, result or scientific synthesis.
This paper mainly addresses another topic and also reports a vitamin D analysis, result or scientific synthesis.
Why this paper is included
Supporting source evidence:
Overall, the data demonstrated that the POCT devices used for HbA1c, Creatinine and Vitamin D testing were comparable to the current reference-standard venous blood sample assays, with strong levels of concordance.
Source: Results. View original paper
Relevance classification does not establish study quality, causation or editorial approval.
Study design Study design awaiting verification.
Study design awaiting verification
The available source does not establish a sufficiently clear study design.
Meta-analysis unclear · Blinding unclear / not reported · Human research reported
How this paper was classified
Automatic source-based classification; awaiting editorial verification. This is not a quality score.
Fifty participants were recruited at Birmingham Dental Hospital and consented for a finger-prick blood spot CityAssay vitamin D test and a venous (control) blood sample.
Source: Methods.
What the study types mean
- Systematic review
- Uses explicit methods to identify, assess and summarise research on a defined question.
- Scoping review
- Maps the extent and characteristics of research on a topic.
- Narrative review
- Discusses existing research; systematic-review methods have not been established by this classification.
- Review / evidence synthesis
- Reviews existing research; the exact review methods need further assessment.
- Randomised controlled trial
- Assigns participants or groups to comparison interventions by chance.
- Non-randomised intervention study
- Examines an intervention without random allocation.
- Cohort study
- Examines exposures and outcomes over time within a defined group.
- Case-control study
- Compares people with an outcome against people without it, examining exposures.
- Cross-sectional study
- Examines a population at a particular point in time.
- Observational study
- Observes exposures and outcomes without assigning the exposure; the subtype is not established here.
- Case report
- Describes an individual case.
- Case series
- Describes a group of individual cases.
- Secondary analysis
- Analyses existing study data; this paper is not automatically another independent trial.
- Laboratory study
- Investigates biological processes in a laboratory setting.
- Qualitative study
- Examines experiences or meanings using qualitative methods.
- Genetic instrumental-variable study
- Uses genetic variation as an instrument to investigate relationships; this is not a randomised treatment trial.
- Study protocol
- Describes planned methods and is not evidence that the planned study has reported results.
- Multiple reported designs
- More than one design is explicitly reported; examine the supporting methods.
- Study design awaiting verification
- The available source does not establish a sufficiently clear study design.
Meta-analysis combines numerical results from studies. Blinding describes who was unaware of treatment allocation. Neither is a separate quality grade. Study design, vitamin D relevance and document format are independent filters.
Abstracts
Concordance of three point of care testing devices with clinical chemistry laboratory standard assays and patient-reported outcomes of blood sampling methods
Study design awaiting verification · BMC Medical Informatics and Decision Making · 2022
Original publication title
Concordance of three point of care testing devices with clinical chemistry laboratory standard assays and patient-reported outcomes of blood sampling methods
© The Author(s) 2022
Original publication · Publisher record · CC BY 4.0 · CC0
Source wording retained. Formatting, section navigation and display-title punctuation adapted for this research record.
At a glance
Abstract: What background does the paper provide?
Point of care testing (POCT) devices have been developed to facilitate immediate results with the potential to aid screening for new disease and enable patients to self-monitor their disease. Non-communicable diseases (NCDs) are the major cause of mortality globally and are increasing in prevalence as the population ages. Allied health care professionals (AHPs) are skilled in undertaking risk assessment and delivering preventative advice, providing opportunities to access large proportions of the population who may not visit their doctor, within non-traditional community settings.
There is evidence of high levels of support from public, patients and health professionals for engaging AHPs in risk-targeted early case detection of certain NCDs. Thus, POCT devices offer a potential alternative to traditional venous blood collection, as novel care pathways for increasing early case detection and access to preventative care. The objectives of this study were to: (i) determine the concordance of the specific POCT devices with laboratory-based standard assays employed within clinical biochemistry laboratories. (ii) compare the sampling experience of both methods via patient-reported experiences.
Source: BackgroundAbstract: What methods does the paper describe?
A prospective, two-centre study was undertaken involving 158 participants who provided informed consent. Venous blood was collected for traditional assays of HbA1c, creatinine/ estimated Glomerular-Filtration-Rate (eGFR) and vitamin-D. Capillary blood was collected by finger prick test and also assayed for the same biochemical indices (Nova StatSensor (creatinine/eGFR); Siemens DCA-Vantage (HbA1C); CityAssays (vitamin-D)). All users were provided with device training. Participants reported any discomfort experienced by each simultaneously applied method (randomised in order) via a 100 mm Visual-Analogue-Scale.
Source: MethodsAbstract: Results
Results for each POCT device and the laboratory standard were analysed by Bland-Altman plots to determine assay concordance. POCT devices demonstrated good concordance with laboratory testing, with at least 95% of all samples being within two standard deviations, for each of the devices tested. The majority of participants reported less discomfort with POCT than venepuncture, with the average reported discomfort being 17/100 mm less for POCT compared to venous blood sample collection on the visual analogue scale.
Source: ResultsAbstract: Conclusions
The POCT devices demonstrated acceptable concordance with laboratory-based assays, and patients reported lower levels of discomfort compared to traditional means of blood collection. This study demonstrates the potential of using these devices as acceptable methods for opportunistic testing of “at-risk” individuals within non-traditional community care settings.
Source: ConclusionsAuthors’ abstract
Background
Point of care testing (POCT) devices have been developed to facilitate immediate results with the potential to aid screening for new disease and enable patients to self-monitor their disease. Non-communicable diseases (NCDs) are the major cause of mortality globally and are increasing in prevalence as the population ages. Allied health care professionals (AHPs) are skilled in undertaking risk assessment and delivering preventative advice, providing opportunities to access large proportions of the population who may not visit their doctor, within non-traditional community settings.
There is evidence of high levels of support from public, patients and health professionals for engaging AHPs in risk-targeted early case detection of certain NCDs. Thus, POCT devices offer a potential alternative to traditional venous blood collection, as novel care pathways for increasing early case detection and access to preventative care. The objectives of this study were to: (i) determine the concordance of the specific POCT devices with laboratory-based standard assays employed within clinical biochemistry laboratories. (ii) compare the sampling experience of both methods via patient-reported experiences.
Methods
A prospective, two-centre study was undertaken involving 158 participants who provided informed consent. Venous blood was collected for traditional assays of HbA1c, creatinine/ estimated Glomerular-Filtration-Rate (eGFR) and vitamin-D. Capillary blood was collected by finger prick test and also assayed for the same biochemical indices (Nova StatSensor (creatinine/eGFR); Siemens DCA-Vantage (HbA1C); CityAssays (vitamin-D)). All users were provided with device training. Participants reported any discomfort experienced by each simultaneously applied method (randomised in order) via a 100 mm Visual-Analogue-Scale.
Results
Results for each POCT device and the laboratory standard were analysed by Bland-Altman plots to determine assay concordance. POCT devices demonstrated good concordance with laboratory testing, with at least 95% of all samples being within two standard deviations, for each of the devices tested. The majority of participants reported less discomfort with POCT than venepuncture, with the average reported discomfort being 17/100 mm less for POCT compared to venous blood sample collection on the visual analogue scale.
Conclusions
The POCT devices demonstrated acceptable concordance with laboratory-based assays, and patients reported lower levels of discomfort compared to traditional means of blood collection. This study demonstrates the potential of using these devices as acceptable methods for opportunistic testing of “at-risk” individuals within non-traditional community care settings.
Publication and source details
Original title: Concordance of three point of care testing devices with clinical chemistry laboratory standard assays and patient-reported outcomes of blood sampling methods
Z. Yonel; K. Kuningas; P. Sharma; M. Dutton; Z. Jalal; P. Cockwell; J. Webber; P. Narendran; T. Dietrich; I. L. C. Chapple
BMC Medical Informatics and Decision Making · 2022
© The Author(s) 2022https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.Oral and Dental Research TrustRepository JATS snapshot; version as supplied
Source retrieved: 2026-11-11T08:57:18.348757+00:00
DOI: 10.1186/s12911-022-01999-z · PMC9493167