Study design and methods

Primary relevance to vitamin D Vitamin D is a stated subject of this paper.

Vitamin D is a stated subject of this paper; the supporting source evidence is shown below.

Why this paper is included

Supporting source evidence:

Figure 2 shows the representative chromatograms of 25(OH)D3, 25(OH)D2, 1,25(OH)2D3 and 24,25(OH)2D3 in a pooled human serum sample after derivatization, comparing the conventional SRM (0.7 Th Q1) vs.

Source: S/N improvement by the narrow-window-isolation (NWI)-SRM. View original paper

Relevance classification does not establish study quality, causation or editorial approval.

Study design Study design awaiting verification.

Study design awaiting verification

The available source does not establish a sufficiently clear study design.

Meta-analysis unclear · Blinding unclear / not reported · Research setting awaiting verification

How this paper was classified

Automatic source-based classification; awaiting editorial verification. This is not a quality score.

View supporting evidence in the original paper

What the study types mean
Systematic review
Uses explicit methods to identify, assess and summarise research on a defined question.
Scoping review
Maps the extent and characteristics of research on a topic.
Narrative review
Discusses existing research; systematic-review methods have not been established by this classification.
Review / evidence synthesis
Reviews existing research; the exact review methods need further assessment.
Randomised controlled trial
Assigns participants or groups to comparison interventions by chance.
Non-randomised intervention study
Examines an intervention without random allocation.
Cohort study
Examines exposures and outcomes over time within a defined group.
Case-control study
Compares people with an outcome against people without it, examining exposures.
Cross-sectional study
Examines a population at a particular point in time.
Observational study
Observes exposures and outcomes without assigning the exposure; the subtype is not established here.
Case report
Describes an individual case.
Case series
Describes a group of individual cases.
Secondary analysis
Analyses existing study data; this paper is not automatically another independent trial.
Laboratory study
Investigates biological processes in a laboratory setting.
Qualitative study
Examines experiences or meanings using qualitative methods.
Genetic instrumental-variable study
Uses genetic variation as an instrument to investigate relationships; this is not a randomised treatment trial.
Study protocol
Describes planned methods and is not evidence that the planned study has reported results.
Multiple reported designs
More than one design is explicitly reported; examine the supporting methods.
Study design awaiting verification
The available source does not establish a sufficiently clear study design.

Meta-analysis combines numerical results from studies. Blinding describes who was unaware of treatment allocation. Neither is a separate quality grade. Study design, vitamin D relevance and document format are independent filters.

Full studies

An ultra-sensitive and high-throughput trapping-micro-LC-MS method for quantification of circulating vitamin D metabolites and application in multiple sclerosis patients

Study design awaiting verification · Scientific Reports · 2024

Original publication title

An ultra-sensitive and high-throughput trapping-micro-LC-MS method for quantification of circulating vitamin D metabolites and application in multiple sclerosis patients

Flora Qu; Ming Zhang; Bianca Weinstock-Guttman; Robert Zivadinov; Jun Qu; Xiaoyu Zhu; Murali Ramanathan

© The Author(s) 2024

Original publication · Publisher record · CC BY 4.0

Source wording retained. Formatting, section navigation and display-title punctuation adapted for this research record.

At a glance

Study background

What background does the paper provide?

Vitamin D (VitD) plays a crucial role in a wide variety of biological processes, such as calcium and bone homeostasis, immunomodulation, and cell differentiation1. Circulating 25-hydroxy VitD (25(OH)D), the major circulating metabolite of VitD, is a key marker for VitD deficiency2. The most important biologically-active metabolite, 1,25-dihydroxyvitamin VitD (1,25(OH)2D), is derived from 25(OH)D3. Another di-hydroxyl metabolite, 24,25-dihydroxyvitamin VitD (24,25(OH)2D), is the most abundant metabolite of 25(OH)D, and is important in studying diseases related to VitD catabolism disorder4,5.

VitD has two main forms, D2 and D3. In humans, while D3 can be obtained through cutaneous synthesis or exogenous sources, D2 is only obtained from specific exogenous sources such as fungus6. In clinical practice, VitD supplementation is used to treat VitD deficiency7–9. While D3 is the main type of VitD supplement on the market, D2 is used substantially, for example, by vegans, due to its non-animal source, and in situations where prescribed high-dosage VitD is necessary to treat certain conditions (e.g., severe osteoporosis, hypoparathyroidism, familial hypocholesterolemia and refractory rickets) because the only FDA-approved high-dosage VitD is in the D2 form10,11.

The equivalency of D2 and D3, and their metabolites, in terms of biological effects, remains controversial6,12,13. For example, a recent blood transcriptomic study reveals that D2 and D3 may have considerably diverse effects on the human immune system 14, while different pharmacokinetic properties and biological effects of the metabolites of D2 vs. D3 have been observed15,16. Distinct gene transcription, protein expression and cellular responses in human preosteoblasts upon treatment of 1,25(OH)2D2 and 1,25(OH)2D3 were found17,18.

Therefore, it is crucial to accurately quantify the metabolites of D2 and D3 in order to comprehensively understand the effects of VitD.

Nonetheless, although 25(OH)D3 can be readily quantified due to its relatively high abundance in circulation, other VitD metabolites, such as the di-hydroxyl-metabolites, are challenging to be quantified with sufficient sensitivity and high accuracy because of their low circulating concentrations19. Currently, immunoassays account for more than 90% of the clinical analyses of VitD metabolites, primarily owing to their high throughput20. However, immunoassay methods may be susceptible to problems associated with suboptimal specificity, accuracy and precision21,22.

As a promising alternative, LC-MS-based methods can reliably measure D2 and D3 metabolites because of their high selectivity with a molecular-level resolution. Nevertheless, achieving highly sensitive and accurate quantification of VitD metabolites while maintaining high throughput and robustness remains challenging for LC-MS-based methods21,23,24. Although derivatization is largely employed to enhance the ionization efficiency of VitD metabolites, enabling more sensitive LC-MS analysis25, the improvement in sensitivity may not be sufficient for measuring low-abundance VitD metabolites, especially in individuals with specific diseases where serum levels of these metabolites are anticipated to be lower than in healthy subjects26.

To boost the sensitivity, our group has previously described a strategy using a one-step derivatization with a Cookson-type reagent followed by selective solid-phase extraction and micro-flow LC-MS (µLC-MS) for sensitive quantification of VitD metabolites27,28. The drawback of the method is relatively long analytical cycles and suboptimal robustness, necessitating the cleaning of the LC-MS system after approximately every 200 runs, which limits its application in large clinical cohorts.

To achieve high-throughput and robust analysis of key VitD metabolites with ultra-high sensitivity, we developed a new pipeline in this study, which combines a trapping-micro-LC (T-µLC) method with narrow-window-isolation selected-reaction monitoring MS (NWI-SRM) for analysis of VitD metabolites in serum samples after derivatization. Previously, we developed a T-µLC-MS method for targeted quantification of proteins29. In brief, the system consists of a high-flow LC for rapid online trapping of digested peptide samples on a large-capacity trapping column, and a synchronized low-flow µLC-MS for sensitive analysis29.

Here we modified and further developed the T-µLC method for ultra-sensitive, robust and high throughput quantification of VitD metabolites in serum samples after derivatization. Moreover, for SRM MS detection, we found that utilizing a narrower m/z window at 0.2 Th for precursor isolation on Q1, i.e. the NWI-SRM for MS detection, was synergistic with the T-µLC approach, largely decreasing co-eluted interferences from samples and thereby enhancing both the sensitivity and selectivity for VitD metabolites.

The optimized T-µLC-NWI-SRM system showed salient advantages over the conventional LC-MS-based methods: (i) by combining a selective trapping/delivery approach on the LC part and the NWI-SRM on the MS part, the method substantially improved signal-to-noise (S/N) ratio; (ii) the large-I.D. trap allowed a high-capacity loading, enhancing the signals of VitD metabolites; (iii) the trapping/delivery approach prevented hydrophobic/hydrophilic matrix components from entering the LC-MS, affording exceptional robustness; and iv) the rapid sample trapping and the much simplified matrix for µLC separation facilitated high-throughput analysis.

Collectively, this strategy allowed ultra-sensitive, robust, and high-throughput quantification of multiple D2/D3 metabolites, with the LOQs down to 1.0 pg/mL for 1,25(OH)2D3, 5.0 pg/mL for 24,25(OH)2D3, 30 pg/mL for 25(OH)D2 and 25(OH)D3 within a 9-min analytical cycle.

We applied the established strategy to the quantification of four VitD metabolites in serum samples from 218 multiple sclerosis patients, including 25(OH)D2, 25(OH)D3, 1,25(OH)2D3 and 24,25(OH)2D3. It was reported that the onset and progression of multiple sclerosis, a chronic central nervous system disease, were associated with the level of serum VitD metabolites30. Notably, low blood 25(OH)D levels have been widely reported in multiple sclerosis patients under active disease states31,32. Elevated serum 1,25(OH)2D levels are correlated with a decreasing risk of developing multiple sclerosis33, whereas lower concentrations of the 24,25(OH)2D are associated with increased disability in multiple sclerosis patients28.

Therefore, VitD supplementation is often recommended for multiple sclerosis patients34. Accurate measurement of the levels of VitD metabolites in multiple sclerosis patients will facilitate the interpretation of clinical data and guide disease treatment and management. Moreover, examining the correlation of D2 and D3 metabolites in multiple sclerosis patients would provide valuable insights into VitD metabolism in the patient population, but this remains unexplored. With the developed T-µLC-NWI-SRM method, we successfully quantified all metabolites in all samples, with high throughput exceptional sensitivity, accuracy and reproducibility.

Intriguingly, we observed a negative correlation between the levels of 25(OH)D2 and the levels of 25(OH)D3, 1,25(OH)2D3 and 24,25(OH)2D3 in multiple sclerosis patients, an interesting observation meriting further investigation.

Source: Introduction Source: Introduction Source: Introduction Source: Introduction Source: Introduction

Study methods

What methods does the paper describe?

Chemicals and reagents

The standard of 25(OH)D2 was purchased from Sigma Aldrich (St. Louis, MO). Standards of 25(OH)D3 and 1,25(OH)2D3 were obtained from MilliporeSigma (Burlington, MA). The standard of 24,25(OH)2D3 was from MP Biomedicals (Irvine, CA). Stable isotope labeled internal standards (I.S.) (i.e., d6-1,25(OH)2D3, d6-25(OH)D2 and d6-25(OH)D3) were purchased from Medical Isotope (Pelham, NH). 4-phenyl-1,2,4-triazoline-3,5-dione (PTAD), used as a derivatization reagent, was obtained from Sigma Aldrich (St.

Louis, MO), along with bovine serum albumin (BSA). Phosphate buffered saline (PBS) and HPLC grade acetonitrile (ACN) were purchased from Fisher Chemicals (Waltham, MA). Water was deionized by a Milli-Q water purification system (MilliporeSigma, MA). Formic acid (FA) was purchased from Sigma-Aldrich (St. Louis, MO).

Study population

The study enrolled 218 adult patients with clinically definite multiple sclerosis in Jacobs Multiple Sclerosis Center for Treatment and Research, University at Buffalo. Of the patients,184 were diagnosed with relapsing-remitting multiple sclerosis (RRMS) patients and 34 with progressive multiple sclerosis (PMS), which included both secondary progressive multiple sclerosis (SPMS) and primary progressive MS (PPMS). The demographic characteristics of the multiple sclerosis patient cohort is shown in Supplementary Table 3. The study was conducted in accordance with the Declaration of Helsinki, and the protocol was approved the University at Buffalo Health Sciences Institutional Review Board.

Written informed consent was obtained from all patients at enrollment.

Serum sample preparation

Serum samples were stored at – 80 °C in polypropylene micro-centrifuge tubes until been processed. Once thawed, the serum samples were vortexed, and 200 µL of each serum sample was transferred into a new tube, spiked with 5 µL of an I.S. stock solution containing 36 ng/mL d6-1,25(OH)2D3, 216 ng/mL d6-25(OH)D2 and 216 ng/mL d6-25(OH)D3 in ACN. The samples were placed at 4 °C to equilibrate for 30 min.

Proteins were precipitated by adding 1 mL MeOH, followed by rigorous vortex for 2 min. After centrifugation at 20,000 g for 20 min at 4 °C, the supernatants were transferred to another tube and evaporated under nitrogen stream until completely dried. 200 µL of 1 mg/mL PTAD in ACN was added into the residue for derivatization of VitD metabolites. The reaction was carried out at room temperature for 2 h and terminated by adding 800 µL water. The terminated reaction mixture was dried under nitrogen stream with 37 °C water bath, and then reconstituted with 50 µL of 30% ACN containing 1% FA.

The scheme for derivatization reaction is shown in Supplementary Fig. 8.

Trapping-micro-LC-narrow window isolation-selective reaction monitoring (T-μLC-NWI-SRM)

An UltiMate 3000 LC system (containing an SRD-3400 degasser, NCS-3500RS CAP pumps, a high-flow tertiary gradient pump, and a WPS-3000TBRS autosampler with a 250 μL loop) coupled to a TSQ Quantiva triple-quadrupole mass spectrometer via an Ion Max NG ion source with an H-ESI probe and a 34-G narrow-bore spray needle (Thermo Fisher Scientific, CA) was used. Sample trapping was conducted on a C8 trapping column (15 × 2.1 mm, 3.5 μm particle size, 100 Å, Agilent, CA) at a flow rate of 1 mL/min using the tertiary pump.

The high-flow loading mobile phases (MPs) Atrapping and Btrapping were water:ACN:FA 98:2:0.1 (v/v/v, pH 3.0) and water:ACN:FA 5:95:0.1 (v/v/v, pH 3.0), respectively. A micro-flow selector (5-50 μL/min) was used for μLC-MS. The separation column was an in-house packed column (150 × 0.5 mm, 2.2 µm, 120Å) using the packing material of Acclaim™ 120 C18 column (Thermo Fisher, MA) at a flow rate of 25 μL/min. The micro-flow mobile phases Aanalysis and Banalysis were water:ACN:FA 98:2:0.1 (v/v/v, pH 3.0) and water:ACN:FA 5:95:0.1 (v/v/v, pH 3.0), respectively.

A ZDV six-port valve placed in the heated column compartment was utilized to coordinate the operations of the two flow systems, with the separation temperature controlled at 40 °C. At the beginning of the sample delivery from the trap to the column, a 1-min isocratic elution with the initial gradient Banalysis percentage was used to for peak compression. The LC gradient for the T-μLC system is shown in Supplementary Table 2. The total run time was 9 min.

The MS instrument parameters were set as following: spray voltage was at 3.5 kV, the vaporizer temperature at 50 °C, the sheath gas at 8 arb unit, the auxiliary gas at 6 arb unit, and the capillary temperature was maintained at 325 °C. The optimized RF Lens voltages and collision energies were obtained for each target analyte by an on-the-fly orthogonal-array-optimization (OAO) described in our previous publication29. The isolation window for NWI-SRM was set to 0.2 Th for Q1 and 0.7 Th for Q3 for all channels.

The MS transitions of each target are shown in Supplementary Table 1.

Optimization of the T-μLC conditions

A pooled human serum sample was prepared by taking an equal volume of each serum sample from 25 randomly-selected multiple sclerosis patient serums. The pooled human serum sample was processed using the procedures described in the serum sample preparation section. The sample was then used for T-μLC conditions optimization. To determine the optimal Btrapping%, series of different loading conditions (Btrapping% ranging from 10 to 50%) in a step size of 5% were evaluated with a generic delivery condition (i.e., the trap was not switched offline until the end of analysis) to the analytical column.

The trapping column was equilibrated with the next experimental condition at the end of each analysis cycle to prepare for the next analytical cycle. To determine the optimal Banalysis%, similarly, a series of different delivery conditions (i.e., the Banalysis% during the gradient elution to switch the trap off the column) was evaluated in a step size of 1.5% for the trap-switching Banalysis% ranging from 50 to 85%, with a mild loading condition (Btrapping% = 1%).

Finally, the analytical gradient was finetuned to ensure the separation of the target peak from the endogenous interfering peaks in the shortest run time. To evaluate the sensitivity improvement of four VitD metabolites by high capacity loading, the sample prepared with a pooled human serum was injected at different volumes which equals to 5, 15, 30, 60, 120, 180, 240 or 300 μL serum, and each volume was injected 4 times.

Comparison between signal-to-noise (S/N) ratio of with and without selective-trapping/delivery

A pooled human serum sample was experimentally prepared by combining an equal volume of serum from each of the 25 randomly selected multiple sclerosis patients. The conditions for selective trapping/delivery are described above. To measure S/N ratio of each metabolite without selective trapping/delivery, the same sample was injected, and the T-μLC method was as follows: during sample loading (0–0.3 min), the Btrapping% was set to 1% so that the vast majority of the sample components were retained on the trapping column; after switching the trapping column in line with the analytical column at 0.4 min, the trapping column remained in-line till the end of separation cycle, so that all trapped compounds were delivered to the analytical column.

Samples were analysis in triplicate for either condition.

Calibration and assay validation

An I.S. stock solution was prepared in ACN containing 36 ng/mL d6-1,25(OH)2D3, 216 ng/mL d6-25(OH)D2 and 216 ng/mL d6-25(OH)D3. d6-1,25(OH)2D3 was used as the I.S. for both 1,25(OH)2D3 and 24,25(OH)2D3. Calibration standards were prepared at six levels across the linear range (Table 1) with three replicates at each level by spiking known amounts of VitD metabolites into a surrogate matrix comprised of 50 mg/mL BSA in PBS.

For stability investigation, 20 mL of pooled serum from healthy subjects was spiked with 1 ng/mL or 10 ng/mL of each metabolite, respectively. Aliquots of samples were subjected to various evaluations including multiple freeze-thaw cycles (from − 80 to 22 °C), storage at ambient temperature and storage at 4 °C. At each designated intervals, isotope-labeled I.S. was added, and then the sample were stored at − 80 °C until analysis. Samples was processed using the procedures described above and analzyed with the optimized T-μLC-NWI-SRM method.

For the evaluation of accuracy and precision, blank quality control (QC) samples were prepared by spiking VitD metabolites into the surrogate matrix at four concentration levels including LOQ, and fortified QCs were prepared in a pooled human serum at two levels (Table 2). All QC samples were prepared in six replicates at each QC level. Calibration standards and QC samples were processed using the procedures described above. The optimized T-μLC-NWI-SRM method was used for analysis of calibration standards and QC samples.

The data was processed using Skyline v22.2. Calibration curves were constructed by plotting the ratio of peak area of each metabolite to the peak area of corresponding I.S. at each concentration level, and then fitting the data using weighted least-squares linear regression with a 1/X2 weighting factor. Accuracy and precision were evaluated intra-day on the same day and inter-day on three consecutive days. Accuracy was calculated as the relative error percent of the average concentration relative to the nominal concentration of each QC level.

Precision was calculated as the coefficient of variation of the measured concentrations of six replicates of QC samples (Table 2).

For selectivity evaluation in surrogate matrix, BSA from six preparations were tested. Surrogate matrix was processed using the procedures described above and analzyed with the optimized T-μLC-NWI-SRM method. Selectivity in human serum was tested with pooled human serum sample by mixing an equal volume of each serum sample from 25 randomly-selected multiple sclerosis patient serums. The pooled sample was processed using the procedures described above. A modified T-μLC-NWI-SRM method was used to analyzed the processed sample to enable an 45-min extensive separation gradient on µLC.

Method robustness was evaluated by injecting the pooled human serum sample every ~15 injections of others samples. The response ratio to the first analysis of each metabolite was calculated by dividing the peak intensity in each analysis to the peak intensity in the first analysis.

Table 1

Sensitivity, and linearity range for the quantification of the four metabolites.

VitD metabolites1,25(OH)2D324,25(OH)2D325(OH)D325(OH)D2
LOD* pg/mL0.20.51515
LOQ pg/mL1.05.03030
Linear range (pg/mL)1–8005–40,00030–60,00030-60,000

*Defined as S/N=3.

Table 2

The validation data for quantification of the four VitD metabolites in serum.

VitD metabolitesNominal concentration (pg/mL)Intra-day (n=6)Inter-day (n=6)
Calculated concentration [Mean ± SD, (pg/mL)]Relative error (%)% CVCalculated concentration [Mean ± SD, (pg/mL)]Relative error (%)% CV
Blank QCsa
 1,25(OH)2D31.00.95 ± 0.05− 55.31.12 ± 0.07126.3
6.05.9 ± 0.3− 24.66.2 ± 0.538.0
40.042.6 ± 4.5710.638.5 ± 4.4− 411.0
200.0206.0 ± 7.933.9204.7 ± 10.928.7
 24,25(OH)2D35.05.2 ± 0.549.64.9 ± 0.6−211.2
30.031.4 ± 3.1510.030.7 ± 1.926.4
200.0209.2 ± 11.055.3218.2 ± 10.995.0
1000.01046.9 ± 70.237.21044.2± 75.147.2
 25(OH)D330.031.2 ± 2.848.828.9 ± 3.9− 413.4
1000.01020.0 ± 83.128.1960.0 ± 63.426.6
6670.06620.2 ± 290.3− 14.46280.7 ± 500.1− 68.0
33300.033605.3 ± 1190.113.532823.6 ± 515.6− 11.6
 25(OH)D230.030.7 ± 3.6211.629.8 ± 3.0− 110.0
200.0190.7 ± 8.8− 64.6183.5 ± 13.8− 87.5
1330.01244.1 ± 93.9− 67.51271.7 ± 64.4− 55.1
6670.06412.6 ± 210.2− 43.36152.9 ± 229.8− 83.7
Fortified QCsb,c
 1,25(OH)2D316.715.2 ± 0.9− 95.615.5 ± 0.6− 74.1
50.048.5 ± 2.3− 34.746.5 ± 0.3− 70.7
 24,25(OH)2D3250.0275.0 ± 1.7100.6272.5 ± 22.198.1
750.0832.5 ± 44.1115.3742.5 ± 84.6− 111.4
 25(OH)D31670.01603.2 ± 89.8− 45.61586.5 ± 128.5− 58.1
5000.04600.0 ± 317.4− 86.95150.0 ± 170.033.3
 25(OH)D21670.01569.8 ± 168.0− 610.71653.3 ± 77.7− 14.7
5000.04900.0 ± 137.3− 22.84750.0 ± 427.5− 59.0

aPrepared by spiking standard compounds into blank matrix.

bPrepared by spiking standard compounds into a pooled human serum.

cThe endogenous level has been subtracted from the measured value.

Source: Method Source: Method Source: Method Source: Method Source: Method Source: Method Source: Method Source: Method Source: Method

Development of trapping-micro-LC (T-μLC) strategy for ultra-sensitive, robust and high-throughput quantification of VitD metabolites

Results: Development of trapping-micro-LC (T-μLC) strategy for ultra-sensitive, robust and high-throughput quantification of VitD metabolites

As shown in Fig. 1a, the T-μLC system consists of two synchronized LC components: a high-flow LC for fast and large-capacity sample loading and a low-flow LC for sensitive analysis, enabling highly sensitive, robust and high-throughput quantification of VitD metabolites via three steps:

  • (I)

    Selective, rapid trapping at a high-flow rate of 1000 μL/min on the large I.D. trap, while the trap is not engaged with the analytical column. With an optimized composition of loading mobile phase, the trap selectively concentrates the VitD metabolites from samples, while simultaneously removing more hydrophilic matrix components (e.g. salts and polar organic compounds).

  • (II)

    Selective delivery and peak compression. The trap is switched in line with the analytical column (i.e., the μLC column) and then a microflow gradient at 25 μL/min is used to back-flush the trap, which selectively delivers the VitD metabolites into the μLC-MS system. To prevent peak broadening during this process, two strategies were taken: firstly, we used weaker retention of the trap (C8 stationary phase) vs. analytical column (C18 stationary phase) to enable peak compression when target metabolite is delivered from the trap to column; secondly, retrograde trapping and delivery flows prevent targets from traveling the relatively large trap volume and enable rapid delivery of target metabolites to the μLC column.

    These two approaches enable effective focusing the VitD metabolites at the front of the analytical column. As soon as the four VitD metabolites have been completely eluted to the front of the analytical column, the trap, carrying matrix components that are more hydrophobic than the four VitD metabolites, is switched off the analytical column. In combination with selective trapping, this approach prevents hydrophilic and hydrophobic matrix components from entering the μLC-MS system, Therefore, the μLC-MS only analyzes the fraction containing concentrated VitD metabolites, enabling ultra-sensitive, robust and rapid quantification.

  • (III)

    Simultaneous trap cleaning/equilibration and μLC-MS analysis. The trap has been taken offline and is flushed with a loading mobile phase containing a high composition of organic solvent at a rate of 1000 μL/min to remove the trapped hydrophobic matrix components followed by equilibration of the trap. Simultaneously, the μLC-MS is performing a sensitive analysis of the VitD metabolites.

Figure 1

The design and optimization of a trapping-micro-LC-MS (T-µLC-MS) strategy for ultra-sensitive, high-throughput and robust quantification of VitD metabolites. (a) the steps involved in the analysis of the VitD metabolites by T-µLC-MS. (b) the optimization of Btrapping% for selective trapping of the VitD metabolites and (c) the optimization of the Banalysis% for selective delivery of the VitD metabolites from the trap to the µLC column, using a pooled human serum sample spiked with VitD metabolites.

Figure image at the original source — open the original paper. This file is not included in this text import.

For μLC-MS, we selected a 0.5 mm i.d. C18 column for a balance of sensitivity, robustness, and throughput, and a 3.5 μm material which provided optimal robustness, reproducibility, and separation. A flow rate of 25 μL/min was determined as optimal for μLC, considering resolution, sensitivity, speed, back pressure, and system dead time. For trapping, a 2.1 mm i.d. C8 column was chosen to enhance quantitative loading capacity, with an optimal flow rate of 1 mL/min for fast sample loading, matrix component removal, and equilibration.

The key parameters of the procedure were meticulously optimized to maximize sensitivity, reproducibility and robustness. The T-μLC-MS gradients, as wells as procedures for PTAD derivatization of VitD metabolites are described in “Method”. The MS transitions of each target are shown in Supplementary Table 1.

Substantially increased sensitivity by an optimized selective-trapping and -delivery approach

The most important parameters in the system pertain to the conditions required for selective trapping/delivery of the VitD metabolites, which improves quantitative sensitivity and robustness by effectively removing hydrophobic/hydrophilic matrix components from samples. The performance of selective trapping/delivery was determined by two synchronized parameters: (i) the optimal organic mobile phase B% for high-flow-gradient selective trapping (Btrapping%), and (ii) the optimal organic mobile phase B% of the analytical μLC gradient (Banalysis%) for selective delivery of the VitD metabolites to the column; the trap was switched off the analytical column upon reaching the optimal Banalysis%.

Both parameters were experimentally identified.

The objective for optimization of Btrapping% was to identify the maximal Btrapping% that retained the VitD metabolites during the trapping process while maximizing the removal of more hydrophilic matrix components. As shown in Fig. 1b, the maximum Btrapping% for 25(OH)D2, 25(OH)D3,1,25(OH)2D3 and 24,25(OH)2D3 were in the range of 32-35%. By balancing the considerations between efficient removal of hydrophilic components and the robustness of the procedure, a 31.5% Btrapping% was determined optimal for selective trapping.

The optimization of Banalysis% was to identify the minimal Banalysis% that delivered all four VitD metabolites to the analytical column while retaining more hydrophobic matrix components on the trap. It was observed that all four metabolites were completely transferred to the μLC analytical column when the trap was switched offline at a Banalysis% of 60.5% (Fig. 1c). Consequently, the optimal Banalysis% to switch the trap off the analytical loop was established at 61.5%.

The optimized gradient conditions and the time points for switching the trap on and off the μLC-MS system are shown in Supplementary Table 2. After determining the optimal Btrapping% and Banalysis%, the μLC gradient was fine-tuned to ensure the separation of the target peak from the endogenous interfering peaks within a runtime of 9 mins. The retention time for 25(OH)D3, 25(OH)D2,1,25(OH)2D3 and 24,25(OH)2D3 were 7.44 min, 7.44 min, 6.22 min and 5.10 min, respectively.

This selective trapping/delivery approach effectively reduced the matrix components in the injected samples, leading to reduced chemical noise and improved sensitivity for analyzing the four VitD metabolites. Supplementary Fig. 1 shows the comparison of the S/N with and without the optimized selective trapping/delivery strategy (i.e., 31.5% Btrapping% and 61.5% Banalysis%). Clearly, the selective trapping/delivery strategy resulted in a 1.9–6.1fold increase in the S/N for the analysis of the four VitD metabolites.

S/N improvement by the narrow-window-isolation (NWI)-SRM

The NWI-SRM utilizes a narrower m/z window for isolation of precursors at Q1 (i.e., the first quadrupole), which effectively reduces the co-isolated species with close m/z, and thereby could substantially lower the chemical noise with a typically mild loss of signal intensity29. We have previously reported that NWI-SRM at a 0.2 Th resolution at Q1 enhances the S/N for quantification of proteins in biological samples over SRM with a conventional 0.7 Th window29.

Here, we demonstrated that the NWI-SRM strategy markedly improved the sensitivity for quantification of VitD metabolites. Figure 2 shows the representative chromatograms of 25(OH)D3, 25(OH)D2, 1,25(OH)2D3 and 24,25(OH)2D3 in a pooled human serum sample after derivatization, comparing the conventional SRM (0.7 Th Q1) vs. NWI-SRM (0.2 Th Q1). When employing a 0.2 Th Q1 window, the signal response of target analytes decreased by 50-70% compared to using a conventional 0.7 Th window. However, there was a much higher reduction in chemical noise (>90%), leading to overall improvements in signal-to-noise ratio.

As shown in Fig. 2, the NWI-SRM improved the S/N by 2.1-5.6 folds for the four VitD metabolites.

Figure 2

Comparison of the chromatograms of the four Vitamin D metabolites using (a) conventional Selected Reaction Monitoring (SRM) with Q1 0.7 Th and (b) Narrow-Window-Isolation (NWI)-SRM with Q1 0.2 Th in a pooled human serum sample. Signal-to-noise (S/N) ratios and absolute peak intensities in counts per second (cps) are marked for each metabolite.

Figure image at the original source — open the original paper. This file is not included in this text import.

Further improvement of sensitivity by high-capacity sample loading on the large-I.D. trap

Given the effectively minimized chemical noises by the selective trapping/delivery and NWI-SRM approaches, an approach to enhance signal intensities would further increase the S/N of VitD metabolites. We hypothesized a high-capacity loading of samples, facilitated by the large-I.D. trap, may lead to markedly enhanced sensitivity for the VitD metabolites. To test this hypothesis, we investigated the S/N of the VitD metabolites as a function of loading amounts of samples. For all metabolites, the S/N was found to increase linearly along with the increased loading volumes in the range of 1.25-15 μL per injection, which corresponded to 5-60 μL serum sample per injection (Fig. 3).

However, when the injection volume increased to 30 µL, which corresponded to 120 µL serum, the S/N of 25(OH)D3 or 25(OH)D2 continued to rise, while no significant increase in S/N was observed for 1,25(OH)2D3 or 24,25(OH)2D3. We speculate that this is because the loading capacity of the more hydrophilic analytes such as 1,25(OH)2D3 and 24,25(OH)2D3 have been reached at this volume, for reasons specified previously29. By balancing the considerations of sensitivity gains and serum consumption, the injection volume 15 μL, which corresponded to 60-μL serum loading per analysis, was selected.

Such high-capacity loading further enhanced sensitivity by 2-4 folds compared to the typical loading capacity in a conventional μLC-MS (i.e., injection volumes capped at a volume equal to 15-30 μL serum sample).

Figure 3

Enhanced sensitivity of four VitD metabolites achieved through high-capacity loading. Samples (i.e., reconstituted PTAD-derivatized serum extracts after protein precipitation, see “Method”) equivalent to varying volumes of pooled human serum sample were injected. The signal-to-noise (S/N) ratio of each metabolite at each injection volume was measured in 4 replicates using the optimized trapping-μLC-NWI-SRM method. The x-axis refers to the serum volume corresponding to the actual injected volume. The optimal injection volume (15 μL, corresponding to 60 μL serum) is marked in red.

Two-tailed Student’s t-test p-value<0.01: **; p-value<0.001: ***; ns: no significance.

Figure image at the original source — open the original paper. This file is not included in this text import.

Evaluation of method robustness

For large-cohort clinical analysis, maintaining a high method robustness is crucial. Conventional μLC-MS often experiences markedly decreased operational robustness because of the accumulation of detrimental matrix components on the spray needle, leading to severe signal drop and the need of instrument cleaning after the injection of a limited amount of plasma samples29. By comparison, the selective-trapping and delivery approach efficiently prevents these detrimental matrix components from entering the μLC-MS system while selectively delivering target compounds into the μLC-MS system, boosting the operational robustness.

To evaluate the robustness of the T-μLC-MS system, we monitored the signal intensity of a processed pooled human serum sample injected every ~15 runs of project samples. The signal intensity of each VitD metabolite in the processed pooled human serum sample remained within ±25% of these from the first injection (Supplementary Fig. 2). Therefore, the developed T-μLC-MS method can continuously quantify an extensive cohort (i.e., at least 600 samples) of biological samples without the need to interrupt the runs, indicating excellent robustness.

Evaluation of the overall quantitative sensitivity and method validation

We performed a comprehensive validation on the stability of analytes, as well as the selectivity, sensitivity, accuracy and precision of the developed method.

Stability was tested on several conditions, including multiple freeze-thaw cycles from −80 to 22 °C, storage at 4 °C, and storage at ambient temperature. Both 25(OH)D2 and 25(OH)D3 were found to be highly stable in these conditions, which agreed with previous reports27,35. In line with our previous study27, underivatized 1,25(OH)2D3 was stable at 4 °C over 2 h and at −80 °C for 60 days. Underivatized 1,25(OH)2D3 was found stable for 3 freeze-thaw cycles, and its derivatized form was stable in a cooled autosampler (4 °C) for 96 h.

The 24,25(OH)2D3 was found stable under the above conditions.

Selectivity was evaluated by testing surrogate matrix from six preparations to determine whether endogenous peaks interfered at the SRM transitions chosen for each metabolite or I.S.. Supplementary Fig. 3 shows the representative chromatogram of a blank surrogate matrix. No interference at the retention times of each metabolite or I.S. was observed. The selectivity of each metabolite and I.S. in pooled human serum was confirmed by comparing the ratio of peak area in two transitions for each metabolite or I.S. under multiple reaction monitoring (MRM) mode in the established 9-min method versus the ratio obtained with a 45-min extensive gradient separation.

No additional peaks was separated by extensitve gradient separation (Supplementary Fig. 4), and the ratio of peak area in two MRM transitions was found to be consistent between 9-min method and extensive gradient separation (Supplementary Fig. 5), demonstrating the high selectivity in the pooled human serum.

By incorporating the selective trapping/delivery, NWI-SRM and high-capacity loading, the developed strategy showed exceptional quantitative sensitivity. Within a 9-min analytical cycle, the method achieved LOQs of 1.0 pg/mL for 1,25(OH)2D3, 5.0 pg/mL for 24,25(OH)2D3, 30 pg/mL for both 25(OH)D2 and 25(OH)D3 (Table 1). By comparison, in existing methods, LOQs of 60-200 pg/mL for 1,25(OH)2D3 and 40-700 pg/mL for 24,25(OH)2D3 have been achieved by high-flow LC-MS, and LOQs of 60-200 pg/mL for 1,25(OH)2D3 and 40–700 pg/mL for 24,25(OH)2D3 by μLC-MS3,27,36–42.

As illustrated in Fig. 4, the LOQs achieved by T-µLC-NWI-SRM are superior to existing methods using either high-flow-LC-MS or μLC-MS. Method accuracy and precision were evaluated using quality control (QC) samples prepared by spiking standard compounds into both surrogate matrix (referred as the “blank QC”) and a pooled human serum sample (referred as the “fortified QC”) for which the endogenous levels of the analytes were previously measured. Table 2 shows that, at all QC levels, good accuracies with relative error <15% were obtained for all targeted VitD metabolites, and the quantitative errors were within ±15% of the nominal concentration.

No appreciable matrix effect was observed.

Table 1

Sensitivity, and linearity range for the quantification of the four metabolites.

VitD metabolites1,25(OH)2D324,25(OH)2D325(OH)D325(OH)D2
LOD* pg/mL0.20.51515
LOQ pg/mL1.05.03030
Linear range (pg/mL)1–8005–40,00030–60,00030-60,000

*Defined as S/N=3.

Figure 4

Ultra-high sensitivity of the T-μLC-NWI-SRM method in comparison to published methods utilizing either high-flow LC-MS (flow rate > 200 μL/min) or micro-flow LC-MS (flow rate between 1 and 200 μL/min) for quantification of two di-hydroxy VitD metabolites. The numbers above and below the boxes represent the range of reported Limit of Quantifications (LOQs) in literature.

Figure image at the original source — open the original paper. This file is not included in this text import.
Table 2

The validation data for quantification of the four VitD metabolites in serum.

VitD metabolitesNominal concentration (pg/mL)Intra-day (n=6)Inter-day (n=6)
Calculated concentration [Mean ± SD, (pg/mL)]Relative error (%)% CVCalculated concentration [Mean ± SD, (pg/mL)]Relative error (%)% CV
Blank QCsa
 1,25(OH)2D31.00.95 ± 0.05− 55.31.12 ± 0.07126.3
6.05.9 ± 0.3− 24.66.2 ± 0.538.0
40.042.6 ± 4.5710.638.5 ± 4.4− 411.0
200.0206.0 ± 7.933.9204.7 ± 10.928.7
 24,25(OH)2D35.05.2 ± 0.549.64.9 ± 0.6−211.2
30.031.4 ± 3.1510.030.7 ± 1.926.4
200.0209.2 ± 11.055.3218.2 ± 10.995.0
1000.01046.9 ± 70.237.21044.2± 75.147.2
 25(OH)D330.031.2 ± 2.848.828.9 ± 3.9− 413.4
1000.01020.0 ± 83.128.1960.0 ± 63.426.6
6670.06620.2 ± 290.3− 14.46280.7 ± 500.1− 68.0
33300.033605.3 ± 1190.113.532823.6 ± 515.6− 11.6
 25(OH)D230.030.7 ± 3.6211.629.8 ± 3.0− 110.0
200.0190.7 ± 8.8− 64.6183.5 ± 13.8− 87.5
1330.01244.1 ± 93.9− 67.51271.7 ± 64.4− 55.1
6670.06412.6 ± 210.2− 43.36152.9 ± 229.8− 83.7
Fortified QCsb,c
 1,25(OH)2D316.715.2 ± 0.9− 95.615.5 ± 0.6− 74.1
50.048.5 ± 2.3− 34.746.5 ± 0.3− 70.7
 24,25(OH)2D3250.0275.0 ± 1.7100.6272.5 ± 22.198.1
750.0832.5 ± 44.1115.3742.5 ± 84.6− 111.4
 25(OH)D31670.01603.2 ± 89.8− 45.61586.5 ± 128.5− 58.1
5000.04600.0 ± 317.4− 86.95150.0 ± 170.033.3
 25(OH)D21670.01569.8 ± 168.0− 610.71653.3 ± 77.7− 14.7
5000.04900.0 ± 137.3− 22.84750.0 ± 427.5− 59.0

aPrepared by spiking standard compounds into blank matrix.

bPrepared by spiking standard compounds into a pooled human serum.

cThe endogenous level has been subtracted from the measured value.

Taking together, the validation results showed that the T-μLC-NWI-SRM method achieved excellent quantitative selectivity, sensitivity, accuracy, precision as well as high throughput.

Source: Development of trapping-micro-LC (T-μLC) strategy for ultra-sensitive, robust and high-throughput quantification of VitD metabolites Source: Development of trapping-micro-LC (T-μLC) strategy for ultra-sensitive, robust and high-throughput quantification of VitD metabolites Source: Development of trapping-micro-LC (T-μLC) strategy for ultra-sensitive, robust and high-throughput quantification of VitD metabolites Source: Development of trapping-micro-LC (T-μLC) strategy for ultra-sensitive, robust and high-throughput quantification of VitD metabolites Source: Development of trapping-micro-LC (T-μLC) strategy for ultra-sensitive, robust and high-throughput quantification of VitD metabolites Source: Development of trapping-micro-LC (T-μLC) strategy for ultra-sensitive, robust and high-throughput quantification of VitD metabolites Source: Development of trapping-micro-LC (T-μLC) strategy for ultra-sensitive, robust and high-throughput quantification of VitD metabolites

Application of T-μLC-NWI-SRM method for quantification of VitD metabolites in multiple sclerosis patients

Results: Application of T-μLC-NWI-SRM method for quantification of VitD metabolites in multiple sclerosis patients

We employed the developed T-μLC-NWI-SRM method to analyze serum samples from 218 multiple sclerosis patients. The demographic and clinical characteristics of the multiple sclerosis patient cohort are described in Supplementary Table 3. As shown in Supplementary Fig. 6, the LOQ achieved by T-μLC-NWI-SRM method were below the levels of the four VitD metabolites in the majority of the cohort, enabling a comprehensive investigation of VitD metabolites in the multiple sclerosis patients. Table 3 shows that, in the multiple sclerosis patient cohort, the average total 25(OH)D level was 32.68 ng/mL, and the concentrations of 25(OH)D3 and 25(OH)D2 were 25.63±12.17 ng/mL and 7.04±14.41 ng/mL (mean±SD, same below) respectively.

The concentrations of 1,25(OH)2D3 and 24,25(OH)2D3 were 0.027±0.018 ng/mL and 1.02±0.89 ng/mL respectively. Among these patients, 40% (41/218) had serum 25(OH)D2 levels higher than the published baseline 25(OH)D2 level, (i.e., <5 nmol/L or 2.06 ng/mL43).

Table 3

The average and median concentrations of each VitD metabolite in the multiple sclerosis patient cohort.

Mean (ng/mL)SD (ng/mL)Median (ng/mL)
1,25(OH)2D30.0270.0180.025
24,25(OH)2D31.020.890.81
25(OH)D325.6312.1725.51
25(OH)D27.0414.411.17
Total 25(OH)D*32.6813.5630.83

*Total 25(OH)D is calculated as the sum of 25(OH)D3 and 25(OH)D2 for each patient.

Source: Application of T-μLC-NWI-SRM method for quantification of VitD metabolites in multiple sclerosis patients

The observed negative correlation between the levels of 25(OH)D2 and the levels of three D3 metabolites in multiple sclerosis patients

Results: The observed negative correlation between the levels of 25(OH)D2 and the levels of three D3 metabolites in multiple sclerosis patients

Numerous studies have established a link between the low levels of circulating 25(OH)D and an increased risk and prevalence of multiple sclerosis44, with levels tending to decrease further in later stages of the disease45. Therefore, individuals with multiple sclerosis are suggested to take VitD supplements to maintain circulating 25(OH)D levels34. While both D2 and D3 forms are available as VitD supplements, studies found that their metabolites have different pharmacokinetic properties and biological effects15–18.

Moreover, it was speculated that D2 may impact D3 metabolism6. For example, previous studies in healthy population and patients with chronic kidney disease have suggested a negative correlation between circulating levels of 25(OH)D3 and 25(OH)D2, in either observational or D2 intervention studies46,47.

Using the technique developed here, we examined whether such a negative correlation between 25(OH)D2 and 25(OH)D3 may also be present in multiple sclerosis patients. Furthermore, 25(OH)D3 is metabolized into 1,25(OH)2D3 (Supplementary Fig. 7), which exerts immunomodulatory effects and regulates neurogenesis and neuroprotection, potentially contributing to the potential protective effects of D3 in multiple sclerosis48. The 25(OH)D3 is also metabolized into 24,25(OH)2D3 (Supplementary Fig. 7), which was reported to show a strong association with disability scores of multiple sclerosis patients28.

Thus, this study also aims to investigate whether these di-hydroxyl D3 metabolites correlate with 25(OH)D2 in multiple sclerosis patients, a topic that has not been previously explored.

Using the quantitative data from the cohort of multiple sclerosis patients, we conducted nonparametric Spearman correlation analysis and calculated the linear regression slopes between 25(OH)D2 and each D3 metabolite. The results showed negative correlations of the levels of 25(OH)D2 and the levels of 25(OH)D3 (slope=− 0.26, r=− 0.48, p-value<0.0001) in multiple sclerosis patients, consistent with previous studies in non-multiple-sclerosis populations that demonstrated greater D2 intake being associated with lower D3 metabolite levels6,12,13,46,47,49–51. Moreover, we also observed apparent negative correlations between the levels of 25(OH)D2 and both 1,25(OH)2D3 (slope=− 0.22, r=− 0.44, p-value<0.0001) and 24,25(OH)2D3 (slope=− 0.28, r=− 0.51, p-value<0.0001, Fig. 5).

The negative correlation observed between 25(OH)D2 and the D3 metabolites may result from competition for the shared metabolic pathways between D2 and D3, or from altered elimination pathways for VitD metabolites in individuals with multiple sclerosis. Further investigation is warranted to elucidate the underlying mechanisms.

Figure 5

The negative correlation between the levels of 25(OH)D2 and the levels of 25(OH)2D3, 24,25(OH)2D3, and 25(OH)D3 in the serum samples of 218 multiple sclerosis patients.

Figure image at the original source — open the original paper. This file is not included in this text import.

Source: The observed negative correlation between the levels of 25(OH)D2 and the levels of three D3 metabolites in multiple sclerosis patients Source: The observed negative correlation between the levels of 25(OH)D2 and the levels of three D3 metabolites in multiple sclerosis patients Source: The observed negative correlation between the levels of 25(OH)D2 and the levels of three D3 metabolites in multiple sclerosis patients

Discussion

Here we developed a T-µLC-NWI-SRM method to address the growing demand for sensitive, high-throughput, and robust methods to measure multiple VitD metabolites in clinical samples. Despite that high-flow LC-MS (i.e., flow rate > 200 μL/min) is commonly employed for quantification of VitD metabolites in order to achieve a high-throughput analysis37,39,40,52, they often lacks sufficient sensitivity for the di-hydroxyl VitD metabolites. On the other hand, micro-flow LC-MS (i.e., typical flow rates at 10-50 μL/min) provides higher sensitivity, but is frequently limited by low robustness, low throughput, and the requirement of additional sample preparation procedures to maintain operational robustness27,41,53.

The T-µLC-NWI-SRM method has addressed these challenges by: (i) employing a selective-trapping and delivery approach to efficiently remove matrix components, enabling high-capacity sample loading and enhancing sensitivity, throughput, and robustness; (ii) implementing the NWI-SRM strategy, which further improves sensitivity by providing high selectivity. Taken together, the T-µLC-NWI- SRM achieved LOQs that were markedly lower than any existing LC-MS methods, within a 9-minute short analytical cycle and with exceptional robustness across large cohort clinical samples.

Compared to the existing LC-MS-based methods3,27,36–42, the T-μLC-NWI-SRM strategy developed here achieves drastically lower LOQ for the two di-hydroxyl VitD metabolites (Fig. 4). This ultra-high sensitivity is accomplished through the combination of high-capacity sample loading and selective trapping/delivery on the large I.D. trap, as well as the selective NWI-SRM detection. The T-μLC-NWI-SRM strategy also enhances throughput by providing online sample clean-up, eliminating the needs for solid-phase extraction; additionally, the selective trapping/delivery approach eliminates most of the matrix components so that μLC-MS system can focus on the analysis of the target VitD metabolites, which enables a short analytical cycle of only 9 min.

Finally, the selective trapping/delivery approach also enhances method robustness by preventing detrimental matrix components from entering the µLC-MS system, as evidenced by the absence of perceivable signal loss after over 600 injections of biological samples (Supplementary Fig. 2). Overall, this strategy enables ultra-sensitive, high-throughput, and robust quantification of multiple D2/D3 metabolites in serum samples.

Previous studies reported declined 25(OH)D3 concentrations in both healthy individuals and patients with chronic kidney diseases taking D2 supplementation6,12,46,47. However, the relationship between 25(OH)D2 and the di-hydroxylated VitD metabolites that are critical to the functions of VitD, has been rarely investigated, likely due to technical limitations in sensitive quantification of these metabolites6,46,47. Furthermore, while previous research has suggested a potential interplay between multiple sclerosis and various VitD metabolites, the levels of di-hydroxyl VitD metabolites have not been adequately surveyed in multiple sclerosis patients54.

Here, we measured VitD metabolites in over 200 multiple sclerosis patients and observed a negative correlation between 25(OH)D2 levels and the levels of D3 metabolites, including 25(OH)D3, 1,25(OH)2D3, and 24,25(OH)2D3 in the patient cohorts. The observation may be attributed to the competitive binding of D2 to the saturable VitD 25-hydroxylase enzyme shared by D2 and D355, leading to decreased 25(OH)D3 levels. As 25(OH)D3 is a precursor to both 1,25(OH)2D3 and 24,25(OH)2D356,57, the concentrations of these di-hydroxylated metabolites may also be negatively impacted by D2 ingestion.

This observation could prove valuable in managing multiple sclerosis. One limitation of our data interpretation is the lack of information on the type and dosage of VitD supplements in the patient cohort. However, our observational results are consistent with previous interventional studies that reveals the negative association between 25(OH)D2 and 25(OH)D36,12,46,47.

In summary, we established T-μLC-NWI-SRM strategy allowing sensitive, high-throughput and robust measurement of multiple VitD metabolites in serum samples pretreated by derivatization, which is valuable in large-cohort clinical studies. By applying this method to a large-cohort of serum samples from multiple sclerosis patients, we gained valuable insights into VitD metabolism in multiple sclerosis patients, which facilitates improved treatment and management of the disease.

Source: Discussion Source: Discussion Source: Discussion Source: Discussion

Results

Development of trapping-micro-LC (T-μLC) strategy for ultra-sensitive, robust and high-throughput quantification of VitD metabolites

As shown in Fig. 1a, the T-μLC system consists of two synchronized LC components: a high-flow LC for fast and large-capacity sample loading and a low-flow LC for sensitive analysis, enabling highly sensitive, robust and high-throughput quantification of VitD metabolites via three steps:

  • (I)

    Selective, rapid trapping at a high-flow rate of 1000 μL/min on the large I.D. trap, while the trap is not engaged with the analytical column. With an optimized composition of loading mobile phase, the trap selectively concentrates the VitD metabolites from samples, while simultaneously removing more hydrophilic matrix components (e.g. salts and polar organic compounds).

  • (II)

    Selective delivery and peak compression. The trap is switched in line with the analytical column (i.e., the μLC column) and then a microflow gradient at 25 μL/min is used to back-flush the trap, which selectively delivers the VitD metabolites into the μLC-MS system. To prevent peak broadening during this process, two strategies were taken: firstly, we used weaker retention of the trap (C8 stationary phase) vs. analytical column (C18 stationary phase) to enable peak compression when target metabolite is delivered from the trap to column; secondly, retrograde trapping and delivery flows prevent targets from traveling the relatively large trap volume and enable rapid delivery of target metabolites to the μLC column.

    These two approaches enable effective focusing the VitD metabolites at the front of the analytical column. As soon as the four VitD metabolites have been completely eluted to the front of the analytical column, the trap, carrying matrix components that are more hydrophobic than the four VitD metabolites, is switched off the analytical column. In combination with selective trapping, this approach prevents hydrophilic and hydrophobic matrix components from entering the μLC-MS system, Therefore, the μLC-MS only analyzes the fraction containing concentrated VitD metabolites, enabling ultra-sensitive, robust and rapid quantification.

  • (III)

    Simultaneous trap cleaning/equilibration and μLC-MS analysis. The trap has been taken offline and is flushed with a loading mobile phase containing a high composition of organic solvent at a rate of 1000 μL/min to remove the trapped hydrophobic matrix components followed by equilibration of the trap. Simultaneously, the μLC-MS is performing a sensitive analysis of the VitD metabolites.

Figure 1

The design and optimization of a trapping-micro-LC-MS (T-µLC-MS) strategy for ultra-sensitive, high-throughput and robust quantification of VitD metabolites. (a) the steps involved in the analysis of the VitD metabolites by T-µLC-MS. (b) the optimization of Btrapping% for selective trapping of the VitD metabolites and (c) the optimization of the Banalysis% for selective delivery of the VitD metabolites from the trap to the µLC column, using a pooled human serum sample spiked with VitD metabolites.

Figure image at the original source — open the original paper. This file is not included in this text import.

For μLC-MS, we selected a 0.5 mm i.d. C18 column for a balance of sensitivity, robustness, and throughput, and a 3.5 μm material which provided optimal robustness, reproducibility, and separation. A flow rate of 25 μL/min was determined as optimal for μLC, considering resolution, sensitivity, speed, back pressure, and system dead time. For trapping, a 2.1 mm i.d. C8 column was chosen to enhance quantitative loading capacity, with an optimal flow rate of 1 mL/min for fast sample loading, matrix component removal, and equilibration.

The key parameters of the procedure were meticulously optimized to maximize sensitivity, reproducibility and robustness. The T-μLC-MS gradients, as wells as procedures for PTAD derivatization of VitD metabolites are described in “Method”. The MS transitions of each target are shown in Supplementary Table 1.

Substantially increased sensitivity by an optimized selective-trapping and -delivery approach

The most important parameters in the system pertain to the conditions required for selective trapping/delivery of the VitD metabolites, which improves quantitative sensitivity and robustness by effectively removing hydrophobic/hydrophilic matrix components from samples. The performance of selective trapping/delivery was determined by two synchronized parameters: (i) the optimal organic mobile phase B% for high-flow-gradient selective trapping (Btrapping%), and (ii) the optimal organic mobile phase B% of the analytical μLC gradient (Banalysis%) for selective delivery of the VitD metabolites to the column; the trap was switched off the analytical column upon reaching the optimal Banalysis%.

Both parameters were experimentally identified.

The objective for optimization of Btrapping% was to identify the maximal Btrapping% that retained the VitD metabolites during the trapping process while maximizing the removal of more hydrophilic matrix components. As shown in Fig. 1b, the maximum Btrapping% for 25(OH)D2, 25(OH)D3,1,25(OH)2D3 and 24,25(OH)2D3 were in the range of 32-35%. By balancing the considerations between efficient removal of hydrophilic components and the robustness of the procedure, a 31.5% Btrapping% was determined optimal for selective trapping.

The optimization of Banalysis% was to identify the minimal Banalysis% that delivered all four VitD metabolites to the analytical column while retaining more hydrophobic matrix components on the trap. It was observed that all four metabolites were completely transferred to the μLC analytical column when the trap was switched offline at a Banalysis% of 60.5% (Fig. 1c). Consequently, the optimal Banalysis% to switch the trap off the analytical loop was established at 61.5%.

The optimized gradient conditions and the time points for switching the trap on and off the μLC-MS system are shown in Supplementary Table 2. After determining the optimal Btrapping% and Banalysis%, the μLC gradient was fine-tuned to ensure the separation of the target peak from the endogenous interfering peaks within a runtime of 9 mins. The retention time for 25(OH)D3, 25(OH)D2,1,25(OH)2D3 and 24,25(OH)2D3 were 7.44 min, 7.44 min, 6.22 min and 5.10 min, respectively.

This selective trapping/delivery approach effectively reduced the matrix components in the injected samples, leading to reduced chemical noise and improved sensitivity for analyzing the four VitD metabolites. Supplementary Fig. 1 shows the comparison of the S/N with and without the optimized selective trapping/delivery strategy (i.e., 31.5% Btrapping% and 61.5% Banalysis%). Clearly, the selective trapping/delivery strategy resulted in a 1.9–6.1fold increase in the S/N for the analysis of the four VitD metabolites.

S/N improvement by the narrow-window-isolation (NWI)-SRM

The NWI-SRM utilizes a narrower m/z window for isolation of precursors at Q1 (i.e., the first quadrupole), which effectively reduces the co-isolated species with close m/z, and thereby could substantially lower the chemical noise with a typically mild loss of signal intensity29. We have previously reported that NWI-SRM at a 0.2 Th resolution at Q1 enhances the S/N for quantification of proteins in biological samples over SRM with a conventional 0.7 Th window29.

Here, we demonstrated that the NWI-SRM strategy markedly improved the sensitivity for quantification of VitD metabolites. Figure 2 shows the representative chromatograms of 25(OH)D3, 25(OH)D2, 1,25(OH)2D3 and 24,25(OH)2D3 in a pooled human serum sample after derivatization, comparing the conventional SRM (0.7 Th Q1) vs. NWI-SRM (0.2 Th Q1). When employing a 0.2 Th Q1 window, the signal response of target analytes decreased by 50-70% compared to using a conventional 0.7 Th window. However, there was a much higher reduction in chemical noise (>90%), leading to overall improvements in signal-to-noise ratio.

As shown in Fig. 2, the NWI-SRM improved the S/N by 2.1-5.6 folds for the four VitD metabolites.

Figure 2

Comparison of the chromatograms of the four Vitamin D metabolites using (a) conventional Selected Reaction Monitoring (SRM) with Q1 0.7 Th and (b) Narrow-Window-Isolation (NWI)-SRM with Q1 0.2 Th in a pooled human serum sample. Signal-to-noise (S/N) ratios and absolute peak intensities in counts per second (cps) are marked for each metabolite.

Figure image at the original source — open the original paper. This file is not included in this text import.

Further improvement of sensitivity by high-capacity sample loading on the large-I.D. trap

Given the effectively minimized chemical noises by the selective trapping/delivery and NWI-SRM approaches, an approach to enhance signal intensities would further increase the S/N of VitD metabolites. We hypothesized a high-capacity loading of samples, facilitated by the large-I.D. trap, may lead to markedly enhanced sensitivity for the VitD metabolites. To test this hypothesis, we investigated the S/N of the VitD metabolites as a function of loading amounts of samples. For all metabolites, the S/N was found to increase linearly along with the increased loading volumes in the range of 1.25-15 μL per injection, which corresponded to 5-60 μL serum sample per injection (Fig. 3).

However, when the injection volume increased to 30 µL, which corresponded to 120 µL serum, the S/N of 25(OH)D3 or 25(OH)D2 continued to rise, while no significant increase in S/N was observed for 1,25(OH)2D3 or 24,25(OH)2D3. We speculate that this is because the loading capacity of the more hydrophilic analytes such as 1,25(OH)2D3 and 24,25(OH)2D3 have been reached at this volume, for reasons specified previously29. By balancing the considerations of sensitivity gains and serum consumption, the injection volume 15 μL, which corresponded to 60-μL serum loading per analysis, was selected.

Such high-capacity loading further enhanced sensitivity by 2-4 folds compared to the typical loading capacity in a conventional μLC-MS (i.e., injection volumes capped at a volume equal to 15-30 μL serum sample).

Figure 3

Enhanced sensitivity of four VitD metabolites achieved through high-capacity loading. Samples (i.e., reconstituted PTAD-derivatized serum extracts after protein precipitation, see “Method”) equivalent to varying volumes of pooled human serum sample were injected. The signal-to-noise (S/N) ratio of each metabolite at each injection volume was measured in 4 replicates using the optimized trapping-μLC-NWI-SRM method. The x-axis refers to the serum volume corresponding to the actual injected volume. The optimal injection volume (15 μL, corresponding to 60 μL serum) is marked in red.

Two-tailed Student’s t-test p-value<0.01: **; p-value<0.001: ***; ns: no significance.

Figure image at the original source — open the original paper. This file is not included in this text import.

Evaluation of method robustness

For large-cohort clinical analysis, maintaining a high method robustness is crucial. Conventional μLC-MS often experiences markedly decreased operational robustness because of the accumulation of detrimental matrix components on the spray needle, leading to severe signal drop and the need of instrument cleaning after the injection of a limited amount of plasma samples29. By comparison, the selective-trapping and delivery approach efficiently prevents these detrimental matrix components from entering the μLC-MS system while selectively delivering target compounds into the μLC-MS system, boosting the operational robustness.

To evaluate the robustness of the T-μLC-MS system, we monitored the signal intensity of a processed pooled human serum sample injected every ~15 runs of project samples. The signal intensity of each VitD metabolite in the processed pooled human serum sample remained within ±25% of these from the first injection (Supplementary Fig. 2). Therefore, the developed T-μLC-MS method can continuously quantify an extensive cohort (i.e., at least 600 samples) of biological samples without the need to interrupt the runs, indicating excellent robustness.

Evaluation of the overall quantitative sensitivity and method validation

We performed a comprehensive validation on the stability of analytes, as well as the selectivity, sensitivity, accuracy and precision of the developed method.

Stability was tested on several conditions, including multiple freeze-thaw cycles from −80 to 22 °C, storage at 4 °C, and storage at ambient temperature. Both 25(OH)D2 and 25(OH)D3 were found to be highly stable in these conditions, which agreed with previous reports27,35. In line with our previous study27, underivatized 1,25(OH)2D3 was stable at 4 °C over 2 h and at −80 °C for 60 days. Underivatized 1,25(OH)2D3 was found stable for 3 freeze-thaw cycles, and its derivatized form was stable in a cooled autosampler (4 °C) for 96 h.

The 24,25(OH)2D3 was found stable under the above conditions.

Selectivity was evaluated by testing surrogate matrix from six preparations to determine whether endogenous peaks interfered at the SRM transitions chosen for each metabolite or I.S.. Supplementary Fig. 3 shows the representative chromatogram of a blank surrogate matrix. No interference at the retention times of each metabolite or I.S. was observed. The selectivity of each metabolite and I.S. in pooled human serum was confirmed by comparing the ratio of peak area in two transitions for each metabolite or I.S. under multiple reaction monitoring (MRM) mode in the established 9-min method versus the ratio obtained with a 45-min extensive gradient separation.

No additional peaks was separated by extensitve gradient separation (Supplementary Fig. 4), and the ratio of peak area in two MRM transitions was found to be consistent between 9-min method and extensive gradient separation (Supplementary Fig. 5), demonstrating the high selectivity in the pooled human serum.

By incorporating the selective trapping/delivery, NWI-SRM and high-capacity loading, the developed strategy showed exceptional quantitative sensitivity. Within a 9-min analytical cycle, the method achieved LOQs of 1.0 pg/mL for 1,25(OH)2D3, 5.0 pg/mL for 24,25(OH)2D3, 30 pg/mL for both 25(OH)D2 and 25(OH)D3 (Table 1). By comparison, in existing methods, LOQs of 60-200 pg/mL for 1,25(OH)2D3 and 40-700 pg/mL for 24,25(OH)2D3 have been achieved by high-flow LC-MS, and LOQs of 60-200 pg/mL for 1,25(OH)2D3 and 40–700 pg/mL for 24,25(OH)2D3 by μLC-MS3,27,36–42.

As illustrated in Fig. 4, the LOQs achieved by T-µLC-NWI-SRM are superior to existing methods using either high-flow-LC-MS or μLC-MS. Method accuracy and precision were evaluated using quality control (QC) samples prepared by spiking standard compounds into both surrogate matrix (referred as the “blank QC”) and a pooled human serum sample (referred as the “fortified QC”) for which the endogenous levels of the analytes were previously measured. Table 2 shows that, at all QC levels, good accuracies with relative error <15% were obtained for all targeted VitD metabolites, and the quantitative errors were within ±15% of the nominal concentration.

No appreciable matrix effect was observed.

Table 1

Sensitivity, and linearity range for the quantification of the four metabolites.

VitD metabolites1,25(OH)2D324,25(OH)2D325(OH)D325(OH)D2
LOD* pg/mL0.20.51515
LOQ pg/mL1.05.03030
Linear range (pg/mL)1–8005–40,00030–60,00030-60,000

*Defined as S/N=3.

Figure 4

Ultra-high sensitivity of the T-μLC-NWI-SRM method in comparison to published methods utilizing either high-flow LC-MS (flow rate > 200 μL/min) or micro-flow LC-MS (flow rate between 1 and 200 μL/min) for quantification of two di-hydroxy VitD metabolites. The numbers above and below the boxes represent the range of reported Limit of Quantifications (LOQs) in literature.

Figure image at the original source — open the original paper. This file is not included in this text import.
Table 2

The validation data for quantification of the four VitD metabolites in serum.

VitD metabolitesNominal concentration (pg/mL)Intra-day (n=6)Inter-day (n=6)
Calculated concentration [Mean ± SD, (pg/mL)]Relative error (%)% CVCalculated concentration [Mean ± SD, (pg/mL)]Relative error (%)% CV
Blank QCsa
 1,25(OH)2D31.00.95 ± 0.05− 55.31.12 ± 0.07126.3
6.05.9 ± 0.3− 24.66.2 ± 0.538.0
40.042.6 ± 4.5710.638.5 ± 4.4− 411.0
200.0206.0 ± 7.933.9204.7 ± 10.928.7
 24,25(OH)2D35.05.2 ± 0.549.64.9 ± 0.6−211.2
30.031.4 ± 3.1510.030.7 ± 1.926.4
200.0209.2 ± 11.055.3218.2 ± 10.995.0
1000.01046.9 ± 70.237.21044.2± 75.147.2
 25(OH)D330.031.2 ± 2.848.828.9 ± 3.9− 413.4
1000.01020.0 ± 83.128.1960.0 ± 63.426.6
6670.06620.2 ± 290.3− 14.46280.7 ± 500.1− 68.0
33300.033605.3 ± 1190.113.532823.6 ± 515.6− 11.6
 25(OH)D230.030.7 ± 3.6211.629.8 ± 3.0− 110.0
200.0190.7 ± 8.8− 64.6183.5 ± 13.8− 87.5
1330.01244.1 ± 93.9− 67.51271.7 ± 64.4− 55.1
6670.06412.6 ± 210.2− 43.36152.9 ± 229.8− 83.7
Fortified QCsb,c
 1,25(OH)2D316.715.2 ± 0.9− 95.615.5 ± 0.6− 74.1
50.048.5 ± 2.3− 34.746.5 ± 0.3− 70.7
 24,25(OH)2D3250.0275.0 ± 1.7100.6272.5 ± 22.198.1
750.0832.5 ± 44.1115.3742.5 ± 84.6− 111.4
 25(OH)D31670.01603.2 ± 89.8− 45.61586.5 ± 128.5− 58.1
5000.04600.0 ± 317.4− 86.95150.0 ± 170.033.3
 25(OH)D21670.01569.8 ± 168.0− 610.71653.3 ± 77.7− 14.7
5000.04900.0 ± 137.3− 22.84750.0 ± 427.5− 59.0

aPrepared by spiking standard compounds into blank matrix.

bPrepared by spiking standard compounds into a pooled human serum.

cThe endogenous level has been subtracted from the measured value.

Taking together, the validation results showed that the T-μLC-NWI-SRM method achieved excellent quantitative selectivity, sensitivity, accuracy, precision as well as high throughput.

Application of T-μLC-NWI-SRM method for quantification of VitD metabolites in multiple sclerosis patients

We employed the developed T-μLC-NWI-SRM method to analyze serum samples from 218 multiple sclerosis patients. The demographic and clinical characteristics of the multiple sclerosis patient cohort are described in Supplementary Table 3. As shown in Supplementary Fig. 6, the LOQ achieved by T-μLC-NWI-SRM method were below the levels of the four VitD metabolites in the majority of the cohort, enabling a comprehensive investigation of VitD metabolites in the multiple sclerosis patients. Table 3 shows that, in the multiple sclerosis patient cohort, the average total 25(OH)D level was 32.68 ng/mL, and the concentrations of 25(OH)D3 and 25(OH)D2 were 25.63±12.17 ng/mL and 7.04±14.41 ng/mL (mean±SD, same below) respectively.

The concentrations of 1,25(OH)2D3 and 24,25(OH)2D3 were 0.027±0.018 ng/mL and 1.02±0.89 ng/mL respectively. Among these patients, 40% (41/218) had serum 25(OH)D2 levels higher than the published baseline 25(OH)D2 level, (i.e., <5 nmol/L or 2.06 ng/mL43).

Table 3

The average and median concentrations of each VitD metabolite in the multiple sclerosis patient cohort.

Mean (ng/mL)SD (ng/mL)Median (ng/mL)
1,25(OH)2D30.0270.0180.025
24,25(OH)2D31.020.890.81
25(OH)D325.6312.1725.51
25(OH)D27.0414.411.17
Total 25(OH)D*32.6813.5630.83

*Total 25(OH)D is calculated as the sum of 25(OH)D3 and 25(OH)D2 for each patient.

The observed negative correlation between the levels of 25(OH)D2 and the levels of three D3 metabolites in multiple sclerosis patients

Numerous studies have established a link between the low levels of circulating 25(OH)D and an increased risk and prevalence of multiple sclerosis44, with levels tending to decrease further in later stages of the disease45. Therefore, individuals with multiple sclerosis are suggested to take VitD supplements to maintain circulating 25(OH)D levels34. While both D2 and D3 forms are available as VitD supplements, studies found that their metabolites have different pharmacokinetic properties and biological effects15–18.

Moreover, it was speculated that D2 may impact D3 metabolism6. For example, previous studies in healthy population and patients with chronic kidney disease have suggested a negative correlation between circulating levels of 25(OH)D3 and 25(OH)D2, in either observational or D2 intervention studies46,47.

Using the technique developed here, we examined whether such a negative correlation between 25(OH)D2 and 25(OH)D3 may also be present in multiple sclerosis patients. Furthermore, 25(OH)D3 is metabolized into 1,25(OH)2D3 (Supplementary Fig. 7), which exerts immunomodulatory effects and regulates neurogenesis and neuroprotection, potentially contributing to the potential protective effects of D3 in multiple sclerosis48. The 25(OH)D3 is also metabolized into 24,25(OH)2D3 (Supplementary Fig. 7), which was reported to show a strong association with disability scores of multiple sclerosis patients28.

Thus, this study also aims to investigate whether these di-hydroxyl D3 metabolites correlate with 25(OH)D2 in multiple sclerosis patients, a topic that has not been previously explored.

Using the quantitative data from the cohort of multiple sclerosis patients, we conducted nonparametric Spearman correlation analysis and calculated the linear regression slopes between 25(OH)D2 and each D3 metabolite. The results showed negative correlations of the levels of 25(OH)D2 and the levels of 25(OH)D3 (slope=− 0.26, r=− 0.48, p-value<0.0001) in multiple sclerosis patients, consistent with previous studies in non-multiple-sclerosis populations that demonstrated greater D2 intake being associated with lower D3 metabolite levels6,12,13,46,47,49–51. Moreover, we also observed apparent negative correlations between the levels of 25(OH)D2 and both 1,25(OH)2D3 (slope=− 0.22, r=− 0.44, p-value<0.0001) and 24,25(OH)2D3 (slope=− 0.28, r=− 0.51, p-value<0.0001, Fig. 5).

The negative correlation observed between 25(OH)D2 and the D3 metabolites may result from competition for the shared metabolic pathways between D2 and D3, or from altered elimination pathways for VitD metabolites in individuals with multiple sclerosis. Further investigation is warranted to elucidate the underlying mechanisms.

Figure 5

The negative correlation between the levels of 25(OH)D2 and the levels of 25(OH)2D3, 24,25(OH)2D3, and 25(OH)D3 in the serum samples of 218 multiple sclerosis patients.

Figure image at the original source — open the original paper. This file is not included in this text import.

Source: Results Source: Results Source: Results
Authors’ abstract

Quantitative analysis of the biologically-active metabolites of vitamin D (VitD), which are crucial in regulating various physiological and pathological processes, is important for clinical investigations. Liquid chromatography-tandem mass spectrometry (LC-MS) has been widely used for this purpose but existing LC-MS methods face challenges in achieving highly sensitive and accurate quantification of low-abundance VitD metabolites while maintaining high throughput and robustness. Here we developed a novel pipeline that combines a trapping-micro-LC-(T-µLC) with narrow-window-isolation selected-reaction monitoring MS(NWI-SRM) for ultra-sensitive, robust and high-throughput quantification of VitD metabolites in serum samples after derivatization.

The selective-trapping and delivery approach efficiently removes matrix components, enabling high-capacity sample loading and enhancing sensitivity, throughput, and robustness. The NWI-SRM further improves the sensitivity by providing high selectivity. The lower limits of quantification (LOQs) achieved were markedly lower than any existing LC-MS methods: 1.0 pg/mL for 1,25(OH)2D3, 5.0 pg/mL for 24,25(OH)2D3, 30 pg/mL for both 25(OH)D2 and 25(OH)D3, all within a 9-min cycle. The method is applied to quantify VitD metabolites from 218 patients with multiple sclerosis.

This study revealed negative correlations(r=− 0.44 to − 0.51) between the levels of 25(OH)D2 and all the three D3 metabolites in multiple sclerosis patients.

Read the full paper

Introduction

Vitamin D (VitD) plays a crucial role in a wide variety of biological processes, such as calcium and bone homeostasis, immunomodulation, and cell differentiation1. Circulating 25-hydroxy VitD (25(OH)D), the major circulating metabolite of VitD, is a key marker for VitD deficiency2. The most important biologically-active metabolite, 1,25-dihydroxyvitamin VitD (1,25(OH)2D), is derived from 25(OH)D3. Another di-hydroxyl metabolite, 24,25-dihydroxyvitamin VitD (24,25(OH)2D), is the most abundant metabolite of 25(OH)D, and is important in studying diseases related to VitD catabolism disorder4,5.

VitD has two main forms, D2 and D3. In humans, while D3 can be obtained through cutaneous synthesis or exogenous sources, D2 is only obtained from specific exogenous sources such as fungus6. In clinical practice, VitD supplementation is used to treat VitD deficiency7–9. While D3 is the main type of VitD supplement on the market, D2 is used substantially, for example, by vegans, due to its non-animal source, and in situations where prescribed high-dosage VitD is necessary to treat certain conditions (e.g., severe osteoporosis, hypoparathyroidism, familial hypocholesterolemia and refractory rickets) because the only FDA-approved high-dosage VitD is in the D2 form10,11.

The equivalency of D2 and D3, and their metabolites, in terms of biological effects, remains controversial6,12,13. For example, a recent blood transcriptomic study reveals that D2 and D3 may have considerably diverse effects on the human immune system 14, while different pharmacokinetic properties and biological effects of the metabolites of D2 vs. D3 have been observed15,16. Distinct gene transcription, protein expression and cellular responses in human preosteoblasts upon treatment of 1,25(OH)2D2 and 1,25(OH)2D3 were found17,18.

Therefore, it is crucial to accurately quantify the metabolites of D2 and D3 in order to comprehensively understand the effects of VitD.

Nonetheless, although 25(OH)D3 can be readily quantified due to its relatively high abundance in circulation, other VitD metabolites, such as the di-hydroxyl-metabolites, are challenging to be quantified with sufficient sensitivity and high accuracy because of their low circulating concentrations19. Currently, immunoassays account for more than 90% of the clinical analyses of VitD metabolites, primarily owing to their high throughput20. However, immunoassay methods may be susceptible to problems associated with suboptimal specificity, accuracy and precision21,22.

As a promising alternative, LC-MS-based methods can reliably measure D2 and D3 metabolites because of their high selectivity with a molecular-level resolution. Nevertheless, achieving highly sensitive and accurate quantification of VitD metabolites while maintaining high throughput and robustness remains challenging for LC-MS-based methods21,23,24. Although derivatization is largely employed to enhance the ionization efficiency of VitD metabolites, enabling more sensitive LC-MS analysis25, the improvement in sensitivity may not be sufficient for measuring low-abundance VitD metabolites, especially in individuals with specific diseases where serum levels of these metabolites are anticipated to be lower than in healthy subjects26.

To boost the sensitivity, our group has previously described a strategy using a one-step derivatization with a Cookson-type reagent followed by selective solid-phase extraction and micro-flow LC-MS (µLC-MS) for sensitive quantification of VitD metabolites27,28. The drawback of the method is relatively long analytical cycles and suboptimal robustness, necessitating the cleaning of the LC-MS system after approximately every 200 runs, which limits its application in large clinical cohorts.

To achieve high-throughput and robust analysis of key VitD metabolites with ultra-high sensitivity, we developed a new pipeline in this study, which combines a trapping-micro-LC (T-µLC) method with narrow-window-isolation selected-reaction monitoring MS (NWI-SRM) for analysis of VitD metabolites in serum samples after derivatization. Previously, we developed a T-µLC-MS method for targeted quantification of proteins29. In brief, the system consists of a high-flow LC for rapid online trapping of digested peptide samples on a large-capacity trapping column, and a synchronized low-flow µLC-MS for sensitive analysis29.

Here we modified and further developed the T-µLC method for ultra-sensitive, robust and high throughput quantification of VitD metabolites in serum samples after derivatization. Moreover, for SRM MS detection, we found that utilizing a narrower m/z window at 0.2 Th for precursor isolation on Q1, i.e. the NWI-SRM for MS detection, was synergistic with the T-µLC approach, largely decreasing co-eluted interferences from samples and thereby enhancing both the sensitivity and selectivity for VitD metabolites.

The optimized T-µLC-NWI-SRM system showed salient advantages over the conventional LC-MS-based methods: (i) by combining a selective trapping/delivery approach on the LC part and the NWI-SRM on the MS part, the method substantially improved signal-to-noise (S/N) ratio; (ii) the large-I.D. trap allowed a high-capacity loading, enhancing the signals of VitD metabolites; (iii) the trapping/delivery approach prevented hydrophobic/hydrophilic matrix components from entering the LC-MS, affording exceptional robustness; and iv) the rapid sample trapping and the much simplified matrix for µLC separation facilitated high-throughput analysis.

Collectively, this strategy allowed ultra-sensitive, robust, and high-throughput quantification of multiple D2/D3 metabolites, with the LOQs down to 1.0 pg/mL for 1,25(OH)2D3, 5.0 pg/mL for 24,25(OH)2D3, 30 pg/mL for 25(OH)D2 and 25(OH)D3 within a 9-min analytical cycle.

We applied the established strategy to the quantification of four VitD metabolites in serum samples from 218 multiple sclerosis patients, including 25(OH)D2, 25(OH)D3, 1,25(OH)2D3 and 24,25(OH)2D3. It was reported that the onset and progression of multiple sclerosis, a chronic central nervous system disease, were associated with the level of serum VitD metabolites30. Notably, low blood 25(OH)D levels have been widely reported in multiple sclerosis patients under active disease states31,32. Elevated serum 1,25(OH)2D levels are correlated with a decreasing risk of developing multiple sclerosis33, whereas lower concentrations of the 24,25(OH)2D are associated with increased disability in multiple sclerosis patients28.

Therefore, VitD supplementation is often recommended for multiple sclerosis patients34. Accurate measurement of the levels of VitD metabolites in multiple sclerosis patients will facilitate the interpretation of clinical data and guide disease treatment and management. Moreover, examining the correlation of D2 and D3 metabolites in multiple sclerosis patients would provide valuable insights into VitD metabolism in the patient population, but this remains unexplored. With the developed T-µLC-NWI-SRM method, we successfully quantified all metabolites in all samples, with high throughput exceptional sensitivity, accuracy and reproducibility.

Intriguingly, we observed a negative correlation between the levels of 25(OH)D2 and the levels of 25(OH)D3, 1,25(OH)2D3 and 24,25(OH)2D3 in multiple sclerosis patients, an interesting observation meriting further investigation.

Results

Development of trapping-micro-LC (T-μLC) strategy for ultra-sensitive, robust and high-throughput quantification of VitD metabolites

As shown in Fig. 1a, the T-μLC system consists of two synchronized LC components: a high-flow LC for fast and large-capacity sample loading and a low-flow LC for sensitive analysis, enabling highly sensitive, robust and high-throughput quantification of VitD metabolites via three steps:

  • (I)

    Selective, rapid trapping at a high-flow rate of 1000 μL/min on the large I.D. trap, while the trap is not engaged with the analytical column. With an optimized composition of loading mobile phase, the trap selectively concentrates the VitD metabolites from samples, while simultaneously removing more hydrophilic matrix components (e.g. salts and polar organic compounds).

  • (II)

    Selective delivery and peak compression. The trap is switched in line with the analytical column (i.e., the μLC column) and then a microflow gradient at 25 μL/min is used to back-flush the trap, which selectively delivers the VitD metabolites into the μLC-MS system. To prevent peak broadening during this process, two strategies were taken: firstly, we used weaker retention of the trap (C8 stationary phase) vs. analytical column (C18 stationary phase) to enable peak compression when target metabolite is delivered from the trap to column; secondly, retrograde trapping and delivery flows prevent targets from traveling the relatively large trap volume and enable rapid delivery of target metabolites to the μLC column.

    These two approaches enable effective focusing the VitD metabolites at the front of the analytical column. As soon as the four VitD metabolites have been completely eluted to the front of the analytical column, the trap, carrying matrix components that are more hydrophobic than the four VitD metabolites, is switched off the analytical column. In combination with selective trapping, this approach prevents hydrophilic and hydrophobic matrix components from entering the μLC-MS system, Therefore, the μLC-MS only analyzes the fraction containing concentrated VitD metabolites, enabling ultra-sensitive, robust and rapid quantification.

  • (III)

    Simultaneous trap cleaning/equilibration and μLC-MS analysis. The trap has been taken offline and is flushed with a loading mobile phase containing a high composition of organic solvent at a rate of 1000 μL/min to remove the trapped hydrophobic matrix components followed by equilibration of the trap. Simultaneously, the μLC-MS is performing a sensitive analysis of the VitD metabolites.

Figure 1

The design and optimization of a trapping-micro-LC-MS (T-µLC-MS) strategy for ultra-sensitive, high-throughput and robust quantification of VitD metabolites. (a) the steps involved in the analysis of the VitD metabolites by T-µLC-MS. (b) the optimization of Btrapping% for selective trapping of the VitD metabolites and (c) the optimization of the Banalysis% for selective delivery of the VitD metabolites from the trap to the µLC column, using a pooled human serum sample spiked with VitD metabolites.

Figure image at the original source — open the original paper. This file is not included in this text import.

For μLC-MS, we selected a 0.5 mm i.d. C18 column for a balance of sensitivity, robustness, and throughput, and a 3.5 μm material which provided optimal robustness, reproducibility, and separation. A flow rate of 25 μL/min was determined as optimal for μLC, considering resolution, sensitivity, speed, back pressure, and system dead time. For trapping, a 2.1 mm i.d. C8 column was chosen to enhance quantitative loading capacity, with an optimal flow rate of 1 mL/min for fast sample loading, matrix component removal, and equilibration.

The key parameters of the procedure were meticulously optimized to maximize sensitivity, reproducibility and robustness. The T-μLC-MS gradients, as wells as procedures for PTAD derivatization of VitD metabolites are described in “Method”. The MS transitions of each target are shown in Supplementary Table 1.

Substantially increased sensitivity by an optimized selective-trapping and -delivery approach

The most important parameters in the system pertain to the conditions required for selective trapping/delivery of the VitD metabolites, which improves quantitative sensitivity and robustness by effectively removing hydrophobic/hydrophilic matrix components from samples. The performance of selective trapping/delivery was determined by two synchronized parameters: (i) the optimal organic mobile phase B% for high-flow-gradient selective trapping (Btrapping%), and (ii) the optimal organic mobile phase B% of the analytical μLC gradient (Banalysis%) for selective delivery of the VitD metabolites to the column; the trap was switched off the analytical column upon reaching the optimal Banalysis%.

Both parameters were experimentally identified.

The objective for optimization of Btrapping% was to identify the maximal Btrapping% that retained the VitD metabolites during the trapping process while maximizing the removal of more hydrophilic matrix components. As shown in Fig. 1b, the maximum Btrapping% for 25(OH)D2, 25(OH)D3,1,25(OH)2D3 and 24,25(OH)2D3 were in the range of 32-35%. By balancing the considerations between efficient removal of hydrophilic components and the robustness of the procedure, a 31.5% Btrapping% was determined optimal for selective trapping.

The optimization of Banalysis% was to identify the minimal Banalysis% that delivered all four VitD metabolites to the analytical column while retaining more hydrophobic matrix components on the trap. It was observed that all four metabolites were completely transferred to the μLC analytical column when the trap was switched offline at a Banalysis% of 60.5% (Fig. 1c). Consequently, the optimal Banalysis% to switch the trap off the analytical loop was established at 61.5%.

The optimized gradient conditions and the time points for switching the trap on and off the μLC-MS system are shown in Supplementary Table 2. After determining the optimal Btrapping% and Banalysis%, the μLC gradient was fine-tuned to ensure the separation of the target peak from the endogenous interfering peaks within a runtime of 9 mins. The retention time for 25(OH)D3, 25(OH)D2,1,25(OH)2D3 and 24,25(OH)2D3 were 7.44 min, 7.44 min, 6.22 min and 5.10 min, respectively.

This selective trapping/delivery approach effectively reduced the matrix components in the injected samples, leading to reduced chemical noise and improved sensitivity for analyzing the four VitD metabolites. Supplementary Fig. 1 shows the comparison of the S/N with and without the optimized selective trapping/delivery strategy (i.e., 31.5% Btrapping% and 61.5% Banalysis%). Clearly, the selective trapping/delivery strategy resulted in a 1.9–6.1fold increase in the S/N for the analysis of the four VitD metabolites.

S/N improvement by the narrow-window-isolation (NWI)-SRM

The NWI-SRM utilizes a narrower m/z window for isolation of precursors at Q1 (i.e., the first quadrupole), which effectively reduces the co-isolated species with close m/z, and thereby could substantially lower the chemical noise with a typically mild loss of signal intensity29. We have previously reported that NWI-SRM at a 0.2 Th resolution at Q1 enhances the S/N for quantification of proteins in biological samples over SRM with a conventional 0.7 Th window29.

Here, we demonstrated that the NWI-SRM strategy markedly improved the sensitivity for quantification of VitD metabolites. Figure 2 shows the representative chromatograms of 25(OH)D3, 25(OH)D2, 1,25(OH)2D3 and 24,25(OH)2D3 in a pooled human serum sample after derivatization, comparing the conventional SRM (0.7 Th Q1) vs. NWI-SRM (0.2 Th Q1). When employing a 0.2 Th Q1 window, the signal response of target analytes decreased by 50-70% compared to using a conventional 0.7 Th window. However, there was a much higher reduction in chemical noise (>90%), leading to overall improvements in signal-to-noise ratio.

As shown in Fig. 2, the NWI-SRM improved the S/N by 2.1-5.6 folds for the four VitD metabolites.

Figure 2

Comparison of the chromatograms of the four Vitamin D metabolites using (a) conventional Selected Reaction Monitoring (SRM) with Q1 0.7 Th and (b) Narrow-Window-Isolation (NWI)-SRM with Q1 0.2 Th in a pooled human serum sample. Signal-to-noise (S/N) ratios and absolute peak intensities in counts per second (cps) are marked for each metabolite.

Figure image at the original source — open the original paper. This file is not included in this text import.

Further improvement of sensitivity by high-capacity sample loading on the large-I.D. trap

Given the effectively minimized chemical noises by the selective trapping/delivery and NWI-SRM approaches, an approach to enhance signal intensities would further increase the S/N of VitD metabolites. We hypothesized a high-capacity loading of samples, facilitated by the large-I.D. trap, may lead to markedly enhanced sensitivity for the VitD metabolites. To test this hypothesis, we investigated the S/N of the VitD metabolites as a function of loading amounts of samples. For all metabolites, the S/N was found to increase linearly along with the increased loading volumes in the range of 1.25-15 μL per injection, which corresponded to 5-60 μL serum sample per injection (Fig. 3).

However, when the injection volume increased to 30 µL, which corresponded to 120 µL serum, the S/N of 25(OH)D3 or 25(OH)D2 continued to rise, while no significant increase in S/N was observed for 1,25(OH)2D3 or 24,25(OH)2D3. We speculate that this is because the loading capacity of the more hydrophilic analytes such as 1,25(OH)2D3 and 24,25(OH)2D3 have been reached at this volume, for reasons specified previously29. By balancing the considerations of sensitivity gains and serum consumption, the injection volume 15 μL, which corresponded to 60-μL serum loading per analysis, was selected.

Such high-capacity loading further enhanced sensitivity by 2-4 folds compared to the typical loading capacity in a conventional μLC-MS (i.e., injection volumes capped at a volume equal to 15-30 μL serum sample).

Figure 3

Enhanced sensitivity of four VitD metabolites achieved through high-capacity loading. Samples (i.e., reconstituted PTAD-derivatized serum extracts after protein precipitation, see “Method”) equivalent to varying volumes of pooled human serum sample were injected. The signal-to-noise (S/N) ratio of each metabolite at each injection volume was measured in 4 replicates using the optimized trapping-μLC-NWI-SRM method. The x-axis refers to the serum volume corresponding to the actual injected volume. The optimal injection volume (15 μL, corresponding to 60 μL serum) is marked in red.

Two-tailed Student’s t-test p-value<0.01: **; p-value<0.001: ***; ns: no significance.

Figure image at the original source — open the original paper. This file is not included in this text import.

Evaluation of method robustness

For large-cohort clinical analysis, maintaining a high method robustness is crucial. Conventional μLC-MS often experiences markedly decreased operational robustness because of the accumulation of detrimental matrix components on the spray needle, leading to severe signal drop and the need of instrument cleaning after the injection of a limited amount of plasma samples29. By comparison, the selective-trapping and delivery approach efficiently prevents these detrimental matrix components from entering the μLC-MS system while selectively delivering target compounds into the μLC-MS system, boosting the operational robustness.

To evaluate the robustness of the T-μLC-MS system, we monitored the signal intensity of a processed pooled human serum sample injected every ~15 runs of project samples. The signal intensity of each VitD metabolite in the processed pooled human serum sample remained within ±25% of these from the first injection (Supplementary Fig. 2). Therefore, the developed T-μLC-MS method can continuously quantify an extensive cohort (i.e., at least 600 samples) of biological samples without the need to interrupt the runs, indicating excellent robustness.

Evaluation of the overall quantitative sensitivity and method validation

We performed a comprehensive validation on the stability of analytes, as well as the selectivity, sensitivity, accuracy and precision of the developed method.

Stability was tested on several conditions, including multiple freeze-thaw cycles from −80 to 22 °C, storage at 4 °C, and storage at ambient temperature. Both 25(OH)D2 and 25(OH)D3 were found to be highly stable in these conditions, which agreed with previous reports27,35. In line with our previous study27, underivatized 1,25(OH)2D3 was stable at 4 °C over 2 h and at −80 °C for 60 days. Underivatized 1,25(OH)2D3 was found stable for 3 freeze-thaw cycles, and its derivatized form was stable in a cooled autosampler (4 °C) for 96 h.

The 24,25(OH)2D3 was found stable under the above conditions.

Selectivity was evaluated by testing surrogate matrix from six preparations to determine whether endogenous peaks interfered at the SRM transitions chosen for each metabolite or I.S.. Supplementary Fig. 3 shows the representative chromatogram of a blank surrogate matrix. No interference at the retention times of each metabolite or I.S. was observed. The selectivity of each metabolite and I.S. in pooled human serum was confirmed by comparing the ratio of peak area in two transitions for each metabolite or I.S. under multiple reaction monitoring (MRM) mode in the established 9-min method versus the ratio obtained with a 45-min extensive gradient separation.

No additional peaks was separated by extensitve gradient separation (Supplementary Fig. 4), and the ratio of peak area in two MRM transitions was found to be consistent between 9-min method and extensive gradient separation (Supplementary Fig. 5), demonstrating the high selectivity in the pooled human serum.

By incorporating the selective trapping/delivery, NWI-SRM and high-capacity loading, the developed strategy showed exceptional quantitative sensitivity. Within a 9-min analytical cycle, the method achieved LOQs of 1.0 pg/mL for 1,25(OH)2D3, 5.0 pg/mL for 24,25(OH)2D3, 30 pg/mL for both 25(OH)D2 and 25(OH)D3 (Table 1). By comparison, in existing methods, LOQs of 60-200 pg/mL for 1,25(OH)2D3 and 40-700 pg/mL for 24,25(OH)2D3 have been achieved by high-flow LC-MS, and LOQs of 60-200 pg/mL for 1,25(OH)2D3 and 40–700 pg/mL for 24,25(OH)2D3 by μLC-MS3,27,36–42.

As illustrated in Fig. 4, the LOQs achieved by T-µLC-NWI-SRM are superior to existing methods using either high-flow-LC-MS or μLC-MS. Method accuracy and precision were evaluated using quality control (QC) samples prepared by spiking standard compounds into both surrogate matrix (referred as the “blank QC”) and a pooled human serum sample (referred as the “fortified QC”) for which the endogenous levels of the analytes were previously measured. Table 2 shows that, at all QC levels, good accuracies with relative error <15% were obtained for all targeted VitD metabolites, and the quantitative errors were within ±15% of the nominal concentration.

No appreciable matrix effect was observed.

Table 1

Sensitivity, and linearity range for the quantification of the four metabolites.

VitD metabolites1,25(OH)2D324,25(OH)2D325(OH)D325(OH)D2
LOD* pg/mL0.20.51515
LOQ pg/mL1.05.03030
Linear range (pg/mL)1–8005–40,00030–60,00030-60,000

*Defined as S/N=3.

Figure 4

Ultra-high sensitivity of the T-μLC-NWI-SRM method in comparison to published methods utilizing either high-flow LC-MS (flow rate > 200 μL/min) or micro-flow LC-MS (flow rate between 1 and 200 μL/min) for quantification of two di-hydroxy VitD metabolites. The numbers above and below the boxes represent the range of reported Limit of Quantifications (LOQs) in literature.

Figure image at the original source — open the original paper. This file is not included in this text import.
Table 2

The validation data for quantification of the four VitD metabolites in serum.

VitD metabolitesNominal concentration (pg/mL)Intra-day (n=6)Inter-day (n=6)
Calculated concentration [Mean ± SD, (pg/mL)]Relative error (%)% CVCalculated concentration [Mean ± SD, (pg/mL)]Relative error (%)% CV
Blank QCsa
 1,25(OH)2D31.00.95 ± 0.05− 55.31.12 ± 0.07126.3
6.05.9 ± 0.3− 24.66.2 ± 0.538.0
40.042.6 ± 4.5710.638.5 ± 4.4− 411.0
200.0206.0 ± 7.933.9204.7 ± 10.928.7
 24,25(OH)2D35.05.2 ± 0.549.64.9 ± 0.6−211.2
30.031.4 ± 3.1510.030.7 ± 1.926.4
200.0209.2 ± 11.055.3218.2 ± 10.995.0
1000.01046.9 ± 70.237.21044.2± 75.147.2
 25(OH)D330.031.2 ± 2.848.828.9 ± 3.9− 413.4
1000.01020.0 ± 83.128.1960.0 ± 63.426.6
6670.06620.2 ± 290.3− 14.46280.7 ± 500.1− 68.0
33300.033605.3 ± 1190.113.532823.6 ± 515.6− 11.6
 25(OH)D230.030.7 ± 3.6211.629.8 ± 3.0− 110.0
200.0190.7 ± 8.8− 64.6183.5 ± 13.8− 87.5
1330.01244.1 ± 93.9− 67.51271.7 ± 64.4− 55.1
6670.06412.6 ± 210.2− 43.36152.9 ± 229.8− 83.7
Fortified QCsb,c
 1,25(OH)2D316.715.2 ± 0.9− 95.615.5 ± 0.6− 74.1
50.048.5 ± 2.3− 34.746.5 ± 0.3− 70.7
 24,25(OH)2D3250.0275.0 ± 1.7100.6272.5 ± 22.198.1
750.0832.5 ± 44.1115.3742.5 ± 84.6− 111.4
 25(OH)D31670.01603.2 ± 89.8− 45.61586.5 ± 128.5− 58.1
5000.04600.0 ± 317.4− 86.95150.0 ± 170.033.3
 25(OH)D21670.01569.8 ± 168.0− 610.71653.3 ± 77.7− 14.7
5000.04900.0 ± 137.3− 22.84750.0 ± 427.5− 59.0

aPrepared by spiking standard compounds into blank matrix.

bPrepared by spiking standard compounds into a pooled human serum.

cThe endogenous level has been subtracted from the measured value.

Taking together, the validation results showed that the T-μLC-NWI-SRM method achieved excellent quantitative selectivity, sensitivity, accuracy, precision as well as high throughput.

Application of T-μLC-NWI-SRM method for quantification of VitD metabolites in multiple sclerosis patients

We employed the developed T-μLC-NWI-SRM method to analyze serum samples from 218 multiple sclerosis patients. The demographic and clinical characteristics of the multiple sclerosis patient cohort are described in Supplementary Table 3. As shown in Supplementary Fig. 6, the LOQ achieved by T-μLC-NWI-SRM method were below the levels of the four VitD metabolites in the majority of the cohort, enabling a comprehensive investigation of VitD metabolites in the multiple sclerosis patients. Table 3 shows that, in the multiple sclerosis patient cohort, the average total 25(OH)D level was 32.68 ng/mL, and the concentrations of 25(OH)D3 and 25(OH)D2 were 25.63±12.17 ng/mL and 7.04±14.41 ng/mL (mean±SD, same below) respectively.

The concentrations of 1,25(OH)2D3 and 24,25(OH)2D3 were 0.027±0.018 ng/mL and 1.02±0.89 ng/mL respectively. Among these patients, 40% (41/218) had serum 25(OH)D2 levels higher than the published baseline 25(OH)D2 level, (i.e., <5 nmol/L or 2.06 ng/mL43).

Table 3

The average and median concentrations of each VitD metabolite in the multiple sclerosis patient cohort.

Mean (ng/mL)SD (ng/mL)Median (ng/mL)
1,25(OH)2D30.0270.0180.025
24,25(OH)2D31.020.890.81
25(OH)D325.6312.1725.51
25(OH)D27.0414.411.17
Total 25(OH)D*32.6813.5630.83

*Total 25(OH)D is calculated as the sum of 25(OH)D3 and 25(OH)D2 for each patient.

The observed negative correlation between the levels of 25(OH)D2 and the levels of three D3 metabolites in multiple sclerosis patients

Numerous studies have established a link between the low levels of circulating 25(OH)D and an increased risk and prevalence of multiple sclerosis44, with levels tending to decrease further in later stages of the disease45. Therefore, individuals with multiple sclerosis are suggested to take VitD supplements to maintain circulating 25(OH)D levels34. While both D2 and D3 forms are available as VitD supplements, studies found that their metabolites have different pharmacokinetic properties and biological effects15–18.

Moreover, it was speculated that D2 may impact D3 metabolism6. For example, previous studies in healthy population and patients with chronic kidney disease have suggested a negative correlation between circulating levels of 25(OH)D3 and 25(OH)D2, in either observational or D2 intervention studies46,47.

Using the technique developed here, we examined whether such a negative correlation between 25(OH)D2 and 25(OH)D3 may also be present in multiple sclerosis patients. Furthermore, 25(OH)D3 is metabolized into 1,25(OH)2D3 (Supplementary Fig. 7), which exerts immunomodulatory effects and regulates neurogenesis and neuroprotection, potentially contributing to the potential protective effects of D3 in multiple sclerosis48. The 25(OH)D3 is also metabolized into 24,25(OH)2D3 (Supplementary Fig. 7), which was reported to show a strong association with disability scores of multiple sclerosis patients28.

Thus, this study also aims to investigate whether these di-hydroxyl D3 metabolites correlate with 25(OH)D2 in multiple sclerosis patients, a topic that has not been previously explored.

Using the quantitative data from the cohort of multiple sclerosis patients, we conducted nonparametric Spearman correlation analysis and calculated the linear regression slopes between 25(OH)D2 and each D3 metabolite. The results showed negative correlations of the levels of 25(OH)D2 and the levels of 25(OH)D3 (slope=− 0.26, r=− 0.48, p-value<0.0001) in multiple sclerosis patients, consistent with previous studies in non-multiple-sclerosis populations that demonstrated greater D2 intake being associated with lower D3 metabolite levels6,12,13,46,47,49–51. Moreover, we also observed apparent negative correlations between the levels of 25(OH)D2 and both 1,25(OH)2D3 (slope=− 0.22, r=− 0.44, p-value<0.0001) and 24,25(OH)2D3 (slope=− 0.28, r=− 0.51, p-value<0.0001, Fig. 5).

The negative correlation observed between 25(OH)D2 and the D3 metabolites may result from competition for the shared metabolic pathways between D2 and D3, or from altered elimination pathways for VitD metabolites in individuals with multiple sclerosis. Further investigation is warranted to elucidate the underlying mechanisms.

Figure 5

The negative correlation between the levels of 25(OH)D2 and the levels of 25(OH)2D3, 24,25(OH)2D3, and 25(OH)D3 in the serum samples of 218 multiple sclerosis patients.

Figure image at the original source — open the original paper. This file is not included in this text import.

Discussion

Here we developed a T-µLC-NWI-SRM method to address the growing demand for sensitive, high-throughput, and robust methods to measure multiple VitD metabolites in clinical samples. Despite that high-flow LC-MS (i.e., flow rate > 200 μL/min) is commonly employed for quantification of VitD metabolites in order to achieve a high-throughput analysis37,39,40,52, they often lacks sufficient sensitivity for the di-hydroxyl VitD metabolites. On the other hand, micro-flow LC-MS (i.e., typical flow rates at 10-50 μL/min) provides higher sensitivity, but is frequently limited by low robustness, low throughput, and the requirement of additional sample preparation procedures to maintain operational robustness27,41,53.

The T-µLC-NWI-SRM method has addressed these challenges by: (i) employing a selective-trapping and delivery approach to efficiently remove matrix components, enabling high-capacity sample loading and enhancing sensitivity, throughput, and robustness; (ii) implementing the NWI-SRM strategy, which further improves sensitivity by providing high selectivity. Taken together, the T-µLC-NWI- SRM achieved LOQs that were markedly lower than any existing LC-MS methods, within a 9-minute short analytical cycle and with exceptional robustness across large cohort clinical samples.

Compared to the existing LC-MS-based methods3,27,36–42, the T-μLC-NWI-SRM strategy developed here achieves drastically lower LOQ for the two di-hydroxyl VitD metabolites (Fig. 4). This ultra-high sensitivity is accomplished through the combination of high-capacity sample loading and selective trapping/delivery on the large I.D. trap, as well as the selective NWI-SRM detection. The T-μLC-NWI-SRM strategy also enhances throughput by providing online sample clean-up, eliminating the needs for solid-phase extraction; additionally, the selective trapping/delivery approach eliminates most of the matrix components so that μLC-MS system can focus on the analysis of the target VitD metabolites, which enables a short analytical cycle of only 9 min.

Finally, the selective trapping/delivery approach also enhances method robustness by preventing detrimental matrix components from entering the µLC-MS system, as evidenced by the absence of perceivable signal loss after over 600 injections of biological samples (Supplementary Fig. 2). Overall, this strategy enables ultra-sensitive, high-throughput, and robust quantification of multiple D2/D3 metabolites in serum samples.

Previous studies reported declined 25(OH)D3 concentrations in both healthy individuals and patients with chronic kidney diseases taking D2 supplementation6,12,46,47. However, the relationship between 25(OH)D2 and the di-hydroxylated VitD metabolites that are critical to the functions of VitD, has been rarely investigated, likely due to technical limitations in sensitive quantification of these metabolites6,46,47. Furthermore, while previous research has suggested a potential interplay between multiple sclerosis and various VitD metabolites, the levels of di-hydroxyl VitD metabolites have not been adequately surveyed in multiple sclerosis patients54.

Here, we measured VitD metabolites in over 200 multiple sclerosis patients and observed a negative correlation between 25(OH)D2 levels and the levels of D3 metabolites, including 25(OH)D3, 1,25(OH)2D3, and 24,25(OH)2D3 in the patient cohorts. The observation may be attributed to the competitive binding of D2 to the saturable VitD 25-hydroxylase enzyme shared by D2 and D355, leading to decreased 25(OH)D3 levels. As 25(OH)D3 is a precursor to both 1,25(OH)2D3 and 24,25(OH)2D356,57, the concentrations of these di-hydroxylated metabolites may also be negatively impacted by D2 ingestion.

This observation could prove valuable in managing multiple sclerosis. One limitation of our data interpretation is the lack of information on the type and dosage of VitD supplements in the patient cohort. However, our observational results are consistent with previous interventional studies that reveals the negative association between 25(OH)D2 and 25(OH)D36,12,46,47.

In summary, we established T-μLC-NWI-SRM strategy allowing sensitive, high-throughput and robust measurement of multiple VitD metabolites in serum samples pretreated by derivatization, which is valuable in large-cohort clinical studies. By applying this method to a large-cohort of serum samples from multiple sclerosis patients, we gained valuable insights into VitD metabolism in multiple sclerosis patients, which facilitates improved treatment and management of the disease.

Method

Chemicals and reagents

The standard of 25(OH)D2 was purchased from Sigma Aldrich (St. Louis, MO). Standards of 25(OH)D3 and 1,25(OH)2D3 were obtained from MilliporeSigma (Burlington, MA). The standard of 24,25(OH)2D3 was from MP Biomedicals (Irvine, CA). Stable isotope labeled internal standards (I.S.) (i.e., d6-1,25(OH)2D3, d6-25(OH)D2 and d6-25(OH)D3) were purchased from Medical Isotope (Pelham, NH). 4-phenyl-1,2,4-triazoline-3,5-dione (PTAD), used as a derivatization reagent, was obtained from Sigma Aldrich (St.

Louis, MO), along with bovine serum albumin (BSA). Phosphate buffered saline (PBS) and HPLC grade acetonitrile (ACN) were purchased from Fisher Chemicals (Waltham, MA). Water was deionized by a Milli-Q water purification system (MilliporeSigma, MA). Formic acid (FA) was purchased from Sigma-Aldrich (St. Louis, MO).

Study population

The study enrolled 218 adult patients with clinically definite multiple sclerosis in Jacobs Multiple Sclerosis Center for Treatment and Research, University at Buffalo. Of the patients,184 were diagnosed with relapsing-remitting multiple sclerosis (RRMS) patients and 34 with progressive multiple sclerosis (PMS), which included both secondary progressive multiple sclerosis (SPMS) and primary progressive MS (PPMS). The demographic characteristics of the multiple sclerosis patient cohort is shown in Supplementary Table 3. The study was conducted in accordance with the Declaration of Helsinki, and the protocol was approved the University at Buffalo Health Sciences Institutional Review Board.

Written informed consent was obtained from all patients at enrollment.

Serum sample preparation

Serum samples were stored at – 80 °C in polypropylene micro-centrifuge tubes until been processed. Once thawed, the serum samples were vortexed, and 200 µL of each serum sample was transferred into a new tube, spiked with 5 µL of an I.S. stock solution containing 36 ng/mL d6-1,25(OH)2D3, 216 ng/mL d6-25(OH)D2 and 216 ng/mL d6-25(OH)D3 in ACN. The samples were placed at 4 °C to equilibrate for 30 min.

Proteins were precipitated by adding 1 mL MeOH, followed by rigorous vortex for 2 min. After centrifugation at 20,000 g for 20 min at 4 °C, the supernatants were transferred to another tube and evaporated under nitrogen stream until completely dried. 200 µL of 1 mg/mL PTAD in ACN was added into the residue for derivatization of VitD metabolites. The reaction was carried out at room temperature for 2 h and terminated by adding 800 µL water. The terminated reaction mixture was dried under nitrogen stream with 37 °C water bath, and then reconstituted with 50 µL of 30% ACN containing 1% FA.

The scheme for derivatization reaction is shown in Supplementary Fig. 8.

Trapping-micro-LC-narrow window isolation-selective reaction monitoring (T-μLC-NWI-SRM)

An UltiMate 3000 LC system (containing an SRD-3400 degasser, NCS-3500RS CAP pumps, a high-flow tertiary gradient pump, and a WPS-3000TBRS autosampler with a 250 μL loop) coupled to a TSQ Quantiva triple-quadrupole mass spectrometer via an Ion Max NG ion source with an H-ESI probe and a 34-G narrow-bore spray needle (Thermo Fisher Scientific, CA) was used. Sample trapping was conducted on a C8 trapping column (15 × 2.1 mm, 3.5 μm particle size, 100 Å, Agilent, CA) at a flow rate of 1 mL/min using the tertiary pump.

The high-flow loading mobile phases (MPs) Atrapping and Btrapping were water:ACN:FA 98:2:0.1 (v/v/v, pH 3.0) and water:ACN:FA 5:95:0.1 (v/v/v, pH 3.0), respectively. A micro-flow selector (5-50 μL/min) was used for μLC-MS. The separation column was an in-house packed column (150 × 0.5 mm, 2.2 µm, 120Å) using the packing material of Acclaim™ 120 C18 column (Thermo Fisher, MA) at a flow rate of 25 μL/min. The micro-flow mobile phases Aanalysis and Banalysis were water:ACN:FA 98:2:0.1 (v/v/v, pH 3.0) and water:ACN:FA 5:95:0.1 (v/v/v, pH 3.0), respectively.

A ZDV six-port valve placed in the heated column compartment was utilized to coordinate the operations of the two flow systems, with the separation temperature controlled at 40 °C. At the beginning of the sample delivery from the trap to the column, a 1-min isocratic elution with the initial gradient Banalysis percentage was used to for peak compression. The LC gradient for the T-μLC system is shown in Supplementary Table 2. The total run time was 9 min.

The MS instrument parameters were set as following: spray voltage was at 3.5 kV, the vaporizer temperature at 50 °C, the sheath gas at 8 arb unit, the auxiliary gas at 6 arb unit, and the capillary temperature was maintained at 325 °C. The optimized RF Lens voltages and collision energies were obtained for each target analyte by an on-the-fly orthogonal-array-optimization (OAO) described in our previous publication29. The isolation window for NWI-SRM was set to 0.2 Th for Q1 and 0.7 Th for Q3 for all channels.

The MS transitions of each target are shown in Supplementary Table 1.

Optimization of the T-μLC conditions

A pooled human serum sample was prepared by taking an equal volume of each serum sample from 25 randomly-selected multiple sclerosis patient serums. The pooled human serum sample was processed using the procedures described in the serum sample preparation section. The sample was then used for T-μLC conditions optimization. To determine the optimal Btrapping%, series of different loading conditions (Btrapping% ranging from 10 to 50%) in a step size of 5% were evaluated with a generic delivery condition (i.e., the trap was not switched offline until the end of analysis) to the analytical column.

The trapping column was equilibrated with the next experimental condition at the end of each analysis cycle to prepare for the next analytical cycle. To determine the optimal Banalysis%, similarly, a series of different delivery conditions (i.e., the Banalysis% during the gradient elution to switch the trap off the column) was evaluated in a step size of 1.5% for the trap-switching Banalysis% ranging from 50 to 85%, with a mild loading condition (Btrapping% = 1%).

Finally, the analytical gradient was finetuned to ensure the separation of the target peak from the endogenous interfering peaks in the shortest run time. To evaluate the sensitivity improvement of four VitD metabolites by high capacity loading, the sample prepared with a pooled human serum was injected at different volumes which equals to 5, 15, 30, 60, 120, 180, 240 or 300 μL serum, and each volume was injected 4 times.

Comparison between signal-to-noise (S/N) ratio of with and without selective-trapping/delivery

A pooled human serum sample was experimentally prepared by combining an equal volume of serum from each of the 25 randomly selected multiple sclerosis patients. The conditions for selective trapping/delivery are described above. To measure S/N ratio of each metabolite without selective trapping/delivery, the same sample was injected, and the T-μLC method was as follows: during sample loading (0–0.3 min), the Btrapping% was set to 1% so that the vast majority of the sample components were retained on the trapping column; after switching the trapping column in line with the analytical column at 0.4 min, the trapping column remained in-line till the end of separation cycle, so that all trapped compounds were delivered to the analytical column.

Samples were analysis in triplicate for either condition.

Calibration and assay validation

An I.S. stock solution was prepared in ACN containing 36 ng/mL d6-1,25(OH)2D3, 216 ng/mL d6-25(OH)D2 and 216 ng/mL d6-25(OH)D3. d6-1,25(OH)2D3 was used as the I.S. for both 1,25(OH)2D3 and 24,25(OH)2D3. Calibration standards were prepared at six levels across the linear range (Table 1) with three replicates at each level by spiking known amounts of VitD metabolites into a surrogate matrix comprised of 50 mg/mL BSA in PBS.

For stability investigation, 20 mL of pooled serum from healthy subjects was spiked with 1 ng/mL or 10 ng/mL of each metabolite, respectively. Aliquots of samples were subjected to various evaluations including multiple freeze-thaw cycles (from − 80 to 22 °C), storage at ambient temperature and storage at 4 °C. At each designated intervals, isotope-labeled I.S. was added, and then the sample were stored at − 80 °C until analysis. Samples was processed using the procedures described above and analzyed with the optimized T-μLC-NWI-SRM method.

For the evaluation of accuracy and precision, blank quality control (QC) samples were prepared by spiking VitD metabolites into the surrogate matrix at four concentration levels including LOQ, and fortified QCs were prepared in a pooled human serum at two levels (Table 2). All QC samples were prepared in six replicates at each QC level. Calibration standards and QC samples were processed using the procedures described above. The optimized T-μLC-NWI-SRM method was used for analysis of calibration standards and QC samples.

The data was processed using Skyline v22.2. Calibration curves were constructed by plotting the ratio of peak area of each metabolite to the peak area of corresponding I.S. at each concentration level, and then fitting the data using weighted least-squares linear regression with a 1/X2 weighting factor. Accuracy and precision were evaluated intra-day on the same day and inter-day on three consecutive days. Accuracy was calculated as the relative error percent of the average concentration relative to the nominal concentration of each QC level.

Precision was calculated as the coefficient of variation of the measured concentrations of six replicates of QC samples (Table 2).

For selectivity evaluation in surrogate matrix, BSA from six preparations were tested. Surrogate matrix was processed using the procedures described above and analzyed with the optimized T-μLC-NWI-SRM method. Selectivity in human serum was tested with pooled human serum sample by mixing an equal volume of each serum sample from 25 randomly-selected multiple sclerosis patient serums. The pooled sample was processed using the procedures described above. A modified T-μLC-NWI-SRM method was used to analyzed the processed sample to enable an 45-min extensive separation gradient on µLC.

Method robustness was evaluated by injecting the pooled human serum sample every ~15 injections of others samples. The response ratio to the first analysis of each metabolite was calculated by dividing the peak intensity in each analysis to the peak intensity in the first analysis.

Supplementary Information

Supplementary Information.

Supplement file at the original source — open the original paper. This file is not included in this text import.

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

The online version contains supplementary material available at 10.1038/s41598-024-55939-0.

Acknowledgements

Dr. Murali Ramanathan received research funding from the National Multiple Sclerosis Society, Department of Defense and National Institute of Neurological Diseases and Stroke. He receives royalty from a self-published textbook.

Author contributions

X.Z. conceptualized and designed the study. B.W.G. and R.Z. recruited the patients and collected the clinical samples. F.Q., M.Z. and X.Z. developed and conducted the experiment. F.Q. and X.Z. perform data analysis. F.Q. and X.Z. wrote and organized the manuscript. M.R. revised the manuscript. J.Q. oversaw the mass spectrometry research and provided facilities.

Data availability

The mass spectrometric datasets are available from the corresponding author (X.Z.) on reasonable request. The clinical data are not publicly available due to privacy and ethical restrictions.

Competing interests

The authors declare no competing interests.

References

1. Aranow C Vitamin D and the immune system J. Investig. Med. 2011 59 881 10.2310/JIM.0b013e31821b8755 21527855 PMC3166406
2. Cheng JB Motola DL Mangelsdorf DJ Russell DW De-orphanization of cytochrome P450 2R1: A microsomal vitamin D 25-hydroxylase J. Biol. Chem. 2003 278 38084 38093 10.1074/jbc.M307028200 12867411 PMC4450819
3. Kowalówka M Główka AK Karaźniewicz-Łada M Kosewski G Clinical significance of analysis of vitamin D status in various diseases Nutrients 2020 12 2788 10.3390/nu12092788 32933052 PMC7551674
4. Bosworth CR The serum 24,25-dihydroxyvitamin D concentration, a marker of vitamin D catabolism, is reduced in chronic kidney disease Kidney Int. 2012 82 693 700 10.1038/ki.2012.193 22648296 PMC3434313
5. Ginsberg C The vitamin D metabolite ratio is associated with changes in bone density and fracture risk in older adults J. Bone Miner. Res. 2021 36 2343 2350 10.1002/jbmr.4426 34423858 PMC8688212
6. Martineau AR Differential effects of oral boluses of vitamin D2 vs. vitamin D3 on vitamin D metabolism: A randomized controlled trial J. Clin. Endocrinol. Metab. 2019 104 5831 5839 10.1210/jc.2019-00207 31199458 PMC6797055
7. Kennel KA Drake MT Hurley DL Vitamin D deficiency in adults: When to test and how to treat Mayo Clin. Proc. 2010 85 752 758 10.4065/mcp.2010.0138 20675513 PMC2912737
8. Amrein K Vitamin D deficiency 2.0: An update on the current status worldwide Eur. J. Clin. Nutr. 2020 74 1498 1513 10.1038/s41430-020-0558-y 31959942 PMC7091696
9. Lincoln MR Schneider R Oh J Vitamin D as disease-modifying therapy for multiple sclerosis? Expert. Rev. Clin. Immunol. 2021 17 691 693 10.1080/1744666X.2021.1915772 33836645
10. Streeten EA Mohtasebi Y Konig M Davidoff L Ryan K Hypoparathyroidism: Less severe hypocalcemia with treatment with vitamin D2 compared with calcitriol J. Clin. Endocrinol. Metab. 2017 102 1505 1510 10.1210/jc.2016-3712 28324108 PMC6283446
11. Dominguez LJ Farruggia M Veronese N Barbagallo M Vitamin D sources, metabolism, and deficiency: Available compounds and guidelines for its treatment Metabolites 2021 11 255 10.3390/metabo11040255 33924215 PMC8074587
12. Armas LA Hollis BW Heaney RP Vitamin D2 is much less effective than vitamin D3 in humans J. Clin. Endocrinol. Metab. 2004 89 5387 5391 10.1210/jc.2004-0360 15531486
13. Binkley N Evaluation of ergocalciferol or cholecalciferol dosing, 1600 IU daily or 50,000 IU monthly in older adults J. Clin. Endocrinol. Metab. 2011 96 981 988 10.1210/jc.2010-0015 21289249 PMC3417158
14. Durrant LR Vitamins D2 and D3 have overlapping but different effects on the human immune system revealed through analysis of the blood transcriptome Front. Immunol. 2022 13 767 10.3389/fimmu.2022.790444 PMC8908317 35281034
15. Hoy DA Ramberg CF Jr Horst RL Evidence that discrimination against ergocalciferol by the chick is the result of enhanced metabolic clearance rates for its mono- and dihydroxylated metabolites J. Nutr. 1988 118 633 638 10.1093/jn/118.5.633 2835464
16. Jones G Byrnes B Palma F Segev D Mazur Y Displacement potency of vitamin D2 analogs in competitive protein-binding assays for 25-hydroxyvitamin D3, 24,25-dihydroxyvitamin D3, and 1,25-dihydroxyvitamin D3 J. Clin. Endocrinol. Metab. 1980 50 773 775 10.1210/jcem-50-4-773 6965943
17. Zarei A Hulley PA Sabokbar A Javaid MK Morovat A 25-Hydroxy-and 1α, 25-dihydroxycholecalciferol have greater potencies than 25-Hydroxy-and 1α, 25-dihydroxyergocalciferol in modulating cultured human and mouse osteoblast activities PLoS One 2016 11 e0165462 10.1371/journal.pone.0165462 27893751 PMC5125576
18. Beckman MJ Horst RL Reinhardt TA Beitz DC Up-regulation of the intestinal 1,25-dihydroxyvitamin D receptor during hypervitaminosis D: A comparison between vitamin D2 and vitamin D3 Biochem. Biophys. Res. Commun. 1990 169 910 915 10.1016/0006-291X(90)91979-3 2163637
19. Dirks NF The when, what & how of measuring vitamin D metabolism in clinical medicine Nutrients 2018 10 482 10.3390/nu10040482 29652819 PMC5946267
20. Rola R Kowalski K Bieńkowski T Studzińska S Improved sample preparation method for fast LC-MS/MS analysis of vitamin D metabolites in serum J. Pharm. Biomed. Anal. 2020 190 113529 10.1016/j.jpba.2020.113529 32841780
21. Altieri B Vitamin D testing: Advantages and limits of the current assays Eur. J. Clin. Nutr. 2020 74 231 247 10.1038/s41430-019-0553-3 31907366
22. Heijboer AC Blankenstein MA Kema IP Buijs MM Accuracy of 6 routine 25-hydroxyvitamin D assays: Influence of vitamin D binding protein concentration Clin. Chem. 2012 58 543 548 10.1373/clinchem.2011.176545 22247500
23. Naik M Kamath S Uppangala S Adiga SK Patil A Vitamin D metabolites and analytical challenges Anal. Methods 2023 10.1039/D2AY01692C 36628933
24. Jin Z Bertholf RL Yi X Advances and challenges in the measurement of 1,25-dihydroxyvitamin D: A comprehensive review Crit. Rev. Clin. Lab. Sci. 2023 60 535 548 10.1080/10408363.2023.2212765 37272827
25. Alexandridou A Schorr P Stokes CS Volmer DA Analysis of vitamin D metabolic markers by mass spectrometry: Recent progress regarding the “gold standard” method and integration into clinical practice Mass Spectrom. Rev. 2023 42 1647 1687 10.1002/mas.21768 34967037
26. Sintzel MB Rametta M Reder AT Vitamin D and multiple sclerosis: A comprehensive review Neurol. Ther. 2018 7 59 85 10.1007/s40120-017-0086-4 29243029 PMC5990512
27. Duan X Ultrasensitive quantification of serum vitamin D metabolites using selective solid-phase extraction coupled to microflow liquid chromatography and isotope-dilution mass spectrometry Anal. Chem. 2010 82 2488 2497 10.1021/ac902869y 20151683
28. Weinstock-Guttman B Vitamin D metabolites are associated with clinical and MRI outcomes in multiple sclerosis patients J. Neurol. Neurosurg. Psychiatry 2011 82 189 195 10.1136/jnnp.2010.227942 21047880
29. Zhang M Sensitive, high-throughput, and robust trapping-micro-LC-MS strategy for the quantification of biomarkers and antibody biotherapeutics Anal. Chem. 2018 90 1870 1880 10.1021/acs.analchem.7b03949 29276835 PMC5960441
30. Wesnes K Low vitamin D, but not tobacco use or high BMI, is associated with long-term disability progression in multiple sclerosis Mult. Scler. Relat. Disord. 2021 50 102801 10.1016/j.msard.2021.102801 33636616
31. Gelfand JM Vitamin D in African Americans with multiple sclerosis Neurology 2011 76 1824 1830 10.1212/WNL.0b013e31821cccf5 21606454 PMC3100123
32. Bäcker-Koduah P Vitamin D and disease severity in multiple sclerosis—Baseline data from the randomized controlled trial (EVIDIMS) Front. Neurol. 2020 10.3389/fneur.2020.00129 32158426 PMC7052055
33. Khafaei M Sadeghi Hajiabadi M Abdolmaleki A Role of 1, 25-dihydroxycholecalciferol in immunological and molecular pathways involved in Multiple Sclerosis CAJMPSI 2021 1 55 66
34. Feige J Vitamin D supplementation in multiple sclerosis: A critical analysis of potentials and threats Nutrients 2020 12 783 10.3390/nu12030783 32188044 PMC7146466
35. Lyu H Simultaneous determination of VD2, VD3, 25(OH)D2, and 25(OH)D3 in human plasma using electrospray LC–MS/MS as well as its application to evaluate VD plasma levels in depressive, schizophrenic patients and healthy individuals Biomed. Chromatogr. 2020 34 e4932 10.1002/bmc.4932 32592191
36. Jenkinson C High throughput LC–MS/MS method for the simultaneous analysis of multiple vitamin D analytes in serum J. Chromatogr. B 2016 1014 56 63 10.1016/j.jchromb.2016.01.049 PMC4801106 26874878
37. Kmieć P Vitamin D status including 3-epi-25(OH)D3 among adult patients with thyroid disorders during summer months Endokrynol. Pol. 2018 69 653 660 10.5603/EP.a2018.0065 30259504
38. Dowling KG Hull G Sundvall J Lamberg-Allardt C Cashman KD Improved accuracy of an tandem liquid chromatography–mass spectrometry method measuring 24R,25-dihydroxyvitamin D3 and 25-hydroxyvitamin D metabolites in serum using unspiked controls and its application to determining cross-reactivity of a chemiluminescent microparticle immunoassay J. Chromatogr. A 2017 1497 102 109 10.1016/j.chroma.2017.03.058 28372841
39. Hu T High throughput UHPLC-MS/MS method for the simultaneous quantification of six vitamin D metabolites: Application for vitamin D determination in patients after liver or kidney transplantation Anal. Methods 2020 12 5591 5600 10.1039/D0AY01088J 33174880
40. Rola R Kowalski K Bieńkowski T Kołodyńska-Goworek A Studzińska S Development of a method for multiple vitamin D metabolite measurements by liquid chromatography coupled with tandem mass spectrometry in dried blood spots Analyst 2019 144 299 309 10.1039/C8AN01422A 30403208
41. Wang Z Simultaneous measurement of plasma vitamin D3 metabolites, including 4β,25-dihydroxyvitamin D3, using liquid chromatography–tandem mass spectrometry Anal. Biochem. 2011 418 126 133 10.1016/j.ab.2011.06.043 21784054 PMC3164754
42. Wan D A new sensitive LC/MS/MS analysis of vitamin D metabolites using a click derivatization reagent, 2-nitrosopyridine J. Lipid Res. 2017 58 798 808 10.1194/jlr.D073536 28148721 PMC5392736
43. Wilson LR Tripkovic L Hart KH Lanham-New SA Vitamin D deficiency as a public health issue: Using vitamin D2 or vitamin D3 in future fortification strategies Proc. Nutr. Soc. 2017 76 392 399 10.1017/S0029665117000349 28347378
44. Zhang Y Causal association of genetically determined circulating vitamin D metabolites and calcium with multiple sclerosis in participants of European descent Eur. J. Clin. Nutr. 2023 10.1038/s41430-023-01260-4 36635366
45. Pierrot-Deseilligny C Souberbielle J-C Contribution of vitamin D insufficiency to the pathogenesis of multiple sclerosis Ther. Adv. Neurol. Disord. 2013 6 81 116 10.1177/1756285612473513 23483715 PMC3582312
46. Swanson CM Higher 25(OH)D2 is associated with lower 25(OH)D3 and 1, 25(OH)2D3 J. Clin. Endocrinol. Metab. 2014 99 2736 2744 10.1210/jc.2014-1069 24828488 PMC4121037
47. Batacchi Z Effects of vitamin D2 supplementation on vitamin D3 metabolism in health and CKD Clin. J. Am. Soc. Nephrol. 2017 12 1498 10.2215/CJN.00530117 28768705 PMC5586563
48. Zorzella-Pezavento SFG Is there a window of opportunity for the therapeutic use of vitamin D in multiple sclerosis? Neural Regen. Res. 2022 17 1945 10.4103/1673-5374.335139 35142671 PMC8848597
49. Tjellesen L Hummer L Christiansen C Rødbro P Serum concentration of vitamin D metabolites during treatment with vitamin D2 and D3 in normal premenopausal women Bone Miner. 1986 1 407 413 2849489
50. Lehmann U Bioavailability of vitamin D2 and D3 in healthy volunteers, a randomized placebo-controlled trial J. Clin. Endocrinol. Metab. 2013 98 4339 4345 10.1210/jc.2012-4287 24001747
51. Hammami MM Yusuf A Differential effects of vitamin D2 and D3 supplements on 25-hydroxyvitamin D level are dose, sex, and time dependent: a randomized controlled trial BMC Endocr. Disord. 2017 17 12 10.1186/s12902-017-0163-9 28231782 PMC5324269
52. Ahmed SA Khojah HMJ Al-Thagfan SS Alahmadi YM Mohammed YA An ultrasensitive UHPLC-ESI-MS/MS method augmented with a controlled microwave derivatization reaction for quantitation of vitamin D3 and its major metabolites in COVID-19 patients Talanta 2022 246 123497 10.1016/j.talanta.2022.123497 35487016 PMC9023092
53. Park H Vitamin D metabolism varies among women in different reproductive states consuming the same intakes of vitamin D and related nutrients J. Nutr. 2016 146 1537 1545 10.3945/jn.116.229971 27335139 PMC4958290
54. Miclea A Bagnoud M Chan A Hoepner R A brief review of the effects of vitamin D on multiple sclerosis Front. Immunol. 2020 10.3389/fimmu.2020.00781 32435244 PMC7218089
55. Kragt J Higher levels of 25-hydroxyvitamin D are associated with a lower incidence of multiple sclerosis only in women Mult. Scler. 2009 15 9 15 10.1177/1352458508095920 18701572
56. Clemens TL Zhou XY Myles M Endres D Lindsay R Serum vitamin D2 and vitamin D3 metabolite concentrations and absorption of vitamin D2 in elderly subjects J. Clin. Endocrinol. Metab. 1986 63 656 660 10.1210/jcem-63-3-656 3488327
57. Berg AH 24,25-Dihydroxyvitamin d3 and vitamin D status of community-dwelling black and white Americans Clin. Chem. 2015 61 877 884 10.1373/clinchem.2015.240051 25922442 PMC4686255
Publication and source details

Original title: An ultra-sensitive and high-throughput trapping-micro-LC-MS method for quantification of circulating vitamin D metabolites and application in multiple sclerosis patients

Flora Qu; Ming Zhang; Bianca Weinstock-Guttman; Robert Zivadinov; Jun Qu; Xiaoyu Zhu; Murali Ramanathan

Scientific Reports · 2024

© The Author(s) 2024https://creativecommons.org/licenses/by/4.0/Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Repository JATS snapshot; version as supplied

Source retrieved: 2026-11-11T05:01:28.796861+00:00

DOI: 10.1038/s41598-024-55939-0 · PMC10918069